Skip to main content

Table 1 Frequency of measurements within the normal reference range depending on platelet count

From: Impact of platelet count on results obtained from multiple electrode platelet aggregometry (Multiplate™)

Platelet count (μl-1) 50 100 150 200 250
ADP 30% (6/20) 35% (9/20) 80% (16/20) 81.3% (13/16) 100% (6/6)
ASPI 30% (6/20) 30% (6/20) 65% (13/20) 68.8% (11/16) 83.3% (5/6)
COL 40% (8/20) 50% (10/20) 70% (14/20) 75% (12/16) 66.7% (4/6)
TRAP 30% (6/20) 50% (10/20) 60% (12/20) 68.8% (11/16) 66.7% (4/6)
overall 32.5% (26/80) 43.8% (35/80)   72% (121/168)  
  1. Percentage of measurements within the normal reference range with the absolute number of measurements given in brackets.