Table 3 Product failures of tests, reagents, controls, calibrators and culture media for infection testing (n = 38).

From: Field safety notices released by manufacturers in cases of failure of products for infection testing: analysis of cases reported to the BfArM between 2005 and 2007

Product failure

All n (%)

Subset with FSCA n (%)

Number of cases

38 (100)

32 (100)

No product failure

2 (5.3)

0 (0)

User error

0 (0)

0 (0)

Root cause not identified

2 (5.3)

2 (6.3)

Product failure identified

34 (89.5)

30 (93.7)

   Material defect

6

5

   Software error

1

1

   Calibration error

2

2

   Electrical error

0

0

   Mechanical error

0

0

   Miss of specification

6

5

   Manufacturing error

8

7

   Incorrect instructions for use

2

2

   Non-microbial contamination

0

0

   Packaging error

1

1

   Microbial contamination

3

3

   Interference by other substances

1

0

   Constructional fault

0

0

   Labelling error

4

4