Table 6 Product failures of analysers and general consumables for infection testing (n = 20).

From: Field safety notices released by manufacturers in cases of failure of products for infection testing: analysis of cases reported to the BfArM between 2005 and 2007

Product failure

All analysers1 n (%)

All general consumables1 n (%)

Number of cases

13 (100)

7 (100)

No product failure

0 (0)

0 (0.0)

User error

1 (7.7)

1 (14.3)

Root cause not identified

0 (0)

0 (0)

Product failure identified

12 (92.3)

6 (85.7)

   Material defect

0

1

   Software error

11

1

   Calibration error

0

0

   Electrical error

0

0

   Mechanical error

0

0

   Miss of specification

0

1

   Manufacturing error

0

1

   Incorrect instructions for use

0

0

   Non-microbial contamination

0

0

   Packaging error

0

0

   Microbial contamination

0

1

   Interference by other substances

0

0

   Constructional fault

1

0

   Labelling error

0

1

  1. 1FSCA were performed in all cases.