Figure 3

From: Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty

Figure 3

a: x-migration: horizontal migration of the study population. One box represents one implant. 105 implants (98.1%) show no significant horizontal migration in contrast to two implants (1.9%, black boxes for implants no 61 and 94) with significant horizontal migration. b: y-migration: vertical migration of the study population. One box represents one implant. 102 implants (95.3%) show no significant vertical migration in contrast to 5 implants (4.7%, black boxes for implants no 16, 27, 38, 61 and 94). c: inclination: cup inclination of the study population. One box represents one implant. 104 implants (97.2%) show no significant change of inclination (> 1.7°) in contrast to .3 implants (2.8%, black boxes for implants no 41, 94 and 61).d: anteversion: cup anteversion of the study population. One box represents one implant. 106 implants (99.1%) show no significant change of anteversion (> 1.7°0) in contrast to 1 implant (0.9%, black box for implant no 94).