Skip to main content

Table 1 Material parameters of investigated BMS (italics) and DES (roman)

From: Direct comparison of coronary bare metal vs. drug-eluting stents: same platform, different mechanics?

Stent type Stent material Strut thickness (µm) Coating Drug References
PRO-Kinetic Energy CoCr L-605 60 proBIO amorphous silicon carbide coating [3]
Orsiro CoCr L-605 60 proBIO amorphous silicon carbide coating/BIOlute bioresorbable Poly-l-Lactide Sirolimus (1.4 µg/mm2) [12]
MULTI-LINK 8 CoCr L-605 81 [4, 13]
XIENCE Xpedition CoCr L-605 89 (including coating) Non-erodible polymer poly n-butyl methacrylate (PBMA)/vinylidene fluoride-co-hexafluoropropylene (PVDF-HFP) Everolimus (100 µg/cm2) [11, 14]
REBEL Monorail PtCr 81 [5]
Promus PREMIER PtCr 81 Non-erodible polymer poly n-butyl methacrylate (PBMA)/vinylidene fluoride-co-hexafluoropropylene (PVDF-HFP) Everolimus (100 µg/cm2) [15]
Integrity CoNi MP35 N 90 [6,7,8]
Resolute integrity CoNi MP35 N 90 BioLinx (PVP, C10, and C19) Zotarolimus (1.6 µg/mm2) [7, 16]
Kaname CoCr L-605 80 [9, 10]
Ultimaster CoCr L-605 80 Poly (dl-lactide-co-caprolactone) Sirolimus (3.9 μg/mm stent length) [17]