Skip to main content

Table 3 Mean displacement and stiffness at the fracture site when using the 1-pin and 2-pin external fixator techniques and the T-POD pelvic binder for stabilization of pelvic injuries

From: The T-pod is as stable as supraacetabular fixation using 1 or 2 Schanz screws in partially unstable pelvic fractures: a biomechanical study

  1-pin 2-pin T-POD
Mean displacement (mm)
 Posterior 1.94 ± 0.64 2.15 ± 0.73 2.27 ± 0.80
 Anterior 2.60 ± 1.35 2.44 ± 0.91 2.46 ± 1.11
Mean stiffness (N/mm)
 Posterior 57.04 ± 17.82 50.76 ± 14.07 49.26 ± 16.36