Table 1 Observed radiation dose over the lungs relative to the prescribed dose

From: Lung sparing and ribcage coverage in total body irradiation delivered by helical tomotherapy

Planned dose

N

Minimum (%)

Maximum (%)

Mean (%)

SD (%)

4 Gy

 D95

5

25.60

79.69

60.97

22.68

 D50

5

51.21

94.86

82.00

17.79

 D5

5

96.01

104.60

101.63

3.62

8 Gy

 D95

15

39.95

79.76

54.77

13.20

 D50

15

53.89

99.00

75.28

13.85

 D5

15

87.17

106.11

100.15

4.78

12 Gy

 D95

15

23.00

64.00

37.44

10.28

 D50

15

38.63

83.00

60.39

9.87

 D5

15

69.78

107.00

96.68

8.67