Skip to main content

Table 2 Observed radiation dose over the ribcage relative to the prescribed dose

From: Lung sparing and ribcage coverage in total body irradiation delivered by helical tomotherapy

Planned dose

N

Minimum (%)

Maximum (%)

Mean (%)

SD (%)

4 Gy

 D95

5

81.93

98.81

91.27

6.16

 D50

5

97.00

102.76

100.42

2.16

 D5

5

103.00

108.03

106.031

2.13

8 Gy

 D95

15

71.04

100.93

89.14

9.32

 D50

15

90.42

106.11

99.81

4.21

 D5

15

102.04

110.00

105.60

2.10

12 Gy

 D95

15

57.32

95.06

78.65

11.59

 D50

15

78.50

105.00

96.43

6.34

 D5

15

100.93

114.00

105.82

2.75