Skip to main content

Table 7 Mean reduction in % of coagulation activity under CO2-shielding gas at hard cavitation at a frequency of 27 kHz (n = 5)

From: Novel method for reduction of virus load in blood plasma by sonication

Time60 min120 min180 min240 min300 min
Reduction of coagulation activity under CO2-shielding gas (%)
Quick0 (± 0)5 (± 2)10 (± 2)17 (± 2)24 (± 2)
aPTT0 (± 0)5 (± 2)11 (± 2)18 (± 3)26 (± 3)
Fibrinogen11 (± 2)23 (± 2)30 (± 2)33 (± 3)35 (± 3)
FIX0 (± 0)0 (± 0)0 (± 0)0 (± 0)0 (± 0)
FVIII23 (± 4)41 (± 5)51 (± 5)58 (± 5)61 (± 5)