In-house 3D-printed surgical guides for osseous lesions of the lower jaw: an experimental study

Abstract

Background

The accuracy of computer-assisted biopsies at the lower jaw was compared to the accuracy of freehand biopsies.

Methods

Patients with a bony lesion of the lower jaw with an indication for biopsy were prospectively enrolled. Two customized bone models per patient were produced using a 3D printer. The models of the lower jaw were fitted into a phantom head model to simulate operation room conditions. Biopsies for the study group were taken by means of surgical guides and freehand biopsies were performed for the control group.

Results

The deviation of the biopsy axes from the planning was significantly less when using templates. It turned out to be 1.3 ± 0.6 mm for the biopsies with a surgical guide and 3.9 ± 1.1 mm for the freehand biopsies.

Conclusions

Surgical guides allow significantly higher accuracy of biopsies. The preliminary results are promising, but clinical evaluation is necessary.

Background

For exact diagnosis of osseous oral-maxillofacial pathologies a biopsy is required [1]. Biopsies of osseous jaw tumors can be challenging: in a series of 48 patients, it is reported that ‘Failure to obtain tumor tissue was the most common reason for incorrect diagnosis’ [2]. Besides, this it is important to preserve anatomical structures such as nerves and the periapical region of the teeth [3, 4]. Very recently, Valdec et al. published a case report about the ‘guided biopsy of osseous pathologies in the jaw bone using a 3D-printed, tooth-supported drilling template’ [4]. The case report shows that 3D-printed templates can be successfully used on patients, but since the method was only used on one patient, there are no data from a case series so far.

Therefore, this is a very newly described method and there are no data in the literature to date that indicate whether template-guided biopsies are more accurate than freehand biopsies in the region of the jaw.

However, there are several systems available that allow template-assisted placement of dental implants [5,6,7,8].

In the field of dental implantation, the literature already shows that the template method was associated with fewer errors [9, 10].

The aim of this study was to evaluate the accuracy of biopsies of the lower jaw with the help of 3D-printed surgical templates in a 3D-printed bone model of the lower jaw. The accuracy of computer-assisted biopsies was compared to the accuracy of freehand biopsies. The geometries of the 3D-printed lower jaws were identical to those of the patients, who each had a lesion at the lower jaw with an indication for biopsy. The null hypothesis (H0) is that there is no difference between the accuracy of computer-assisted biopsies and freehand biopsies of the control group.

Methods

Study design

The CT data of patients with a bony lesion of the lower jaw and with an indication for biopsy were used for this experimental study. The use of the CT data was approved by the local ethics committee.

Planning and 3D printing of bone models of the lower jaw

The DICOM (Digital Imaging and Communications in Medicine) data of the CT scan was then transferred to the image segmentation and 3D model creation software Mimics Innovation Suite (Materialise, Leuven, Belgium). After segmentation, the osseous and dental information was available in STL data format. The biopsy channel was then planned considering important anatomical structures such as nerves and teeth roots (Fig. 1).

Fig. 1
figure 1

Part of a screenshot while planning of the biopsy channel according to CT data

Customized bone models of the patients’ lower jaws were produced by a 3D printer (ZPrinter 650, 3D Systems, Rock Hill, South Carolina, USA) according to the previously generated CAD data (STL) of the patients’ jaw. Therefore, the bone modes of lower jaws used in this study showed identical dental and bone geometries to the patients’ lower jaws.

CT scans before operation/experiments

CT scans (Fig. 2) of the 3D-printed jaws (of the guided group) were performed with a Siemens Somatom Force CT scanner (120 kV, 330 mAs, collimation 64 × 0.6 mm, pitch 0.55, slice 0.75 mm).

Fig. 2
figure 2

CT scans of all models of the study group

Design and 3D printing of surgical templates

The data of the CT scans were transferred to the image segmentation and 3D model creation tool Mimics Innovation Suite (Materialise, Leuven, Belgium). A fusion of the planning data and data of the CT scans before surgery was performed. The surgical template was then designed. The depth of the biopsy channel was determined through a limit stop: the head of the contra angle stopped at the surface of the surgical template.

The STL Data of the surgical template was transferred to the software of the 3D Printer (PreForm Software, Formlabs Inc., Somerville, MA, USA) via an USB flash drive. The template was produced by the 3D printer (Form 2, Formlabs Inc., Somerville, MA, USA) using stereolithography with a class 1 biocompatible resin. The printer has an axis resolution of 0.025 mm and a build volume of 125 × 125 × 165 mm. The printer easily fits on the desktop and has a weight of 8 kg. After printing, all models and guides were available (Fig. 3).

Fig. 3
figure 3

Printed models of the control group on the left side and of the study group on the right side (with surgical guides)

Biopsies with surgical guides

The models of the lower jaw were fitted into a phantom head model which was placed on the operating table (Frasaco P6 phantom head, frasaco GmbH, Tettnang, Germany) to simulate operating room conditions (Fig. 4). With the help of the surgical template the biopsy was taken (Fig. 5) with a trephine drill (XiVE Trephine Drill, inner diameter 3.0 mm, outer diameter 4.2 mm, Dentsply, York, PA, USA). All guided biopsies were performed by the same experienced surgeon.

Fig. 4
figure 4

The models were mounted in a phantom head (the left side of picture shows the lower part of the phantom head). The phantom head with a soft tissue mask and upper body torso was placed on the operating table and covered in usual way

Fig. 5
figure 5

Biopsy in the control group

Freehand biopsies/control group

Again, the models of the lower jaw were fitted into a phantom head model (Frasaco P6 phantom head, frasaco GmbH, Tettnang, Germany). Freehand biopsies were taken (Fig. 6) with a trephine drill (XiVE Trephine Drill, inner diameter 3.0 mm, outer diameter 4.2 mm, Dentsply, York, PA, USA). Freehand biopsies were performed about 1 month after the guided biopsies by the same surgeon as for the guided biopsies.

Fig. 6
figure 6

Biopsy in the freehand group

CT scans after operation/experiments

Again, CT scans the models of the 3D-printed jaws were performed with a Siemens Somatom Force CT scanner (120 kV, 330 mAs, collimation 64 × 0.6 mm, pitch 0.55, slice 0.75 mm).

Evaluation

The postoperative CT scans’ data were segmented and loaded into Mimics Innovation Suite software (Materialise, Leuven, Belgium). The true biopsy channel was determined (Fig. 7). A fusion of these postoperative data with the preoperative planning data was performed (Fig. 8). The deviation of the planned biopsy channels to channels drilled in reality was determined by measuring the maximum distance in mm between the axes. Measuring the maximum distance in mm is in accordance with the ISO 1101 standard [11] and has been previously used for determining the accuracy of tumor operations [12, 13]. The angle between the actual biopsy axis and the planned biopsy axis was also determined as well as the depth of the biopsy channels which was compared to the planned depth (Fig. 9).

Fig. 7
figure 7

Part of a screenshot while determination of the biopsy channel

Fig. 8
figure 8

Part of a screenshot after fusion of postoperative data with preoperative planning data for measuring the maximum deviation of axes, of biopsy depth and the angle between the axes

Fig. 9
figure 9

Box plot of the maximum deviation of axes for the guided and the freehand group. The boxes show the range from the first to the third quartile (*p = 0.0002)

Statistics

Values are given as mean values (arithmetic mean) and standard deviations which were calculated with the software IBM SPSS Statistics 22 (IBM, Armonk, NY, USA). The proof of normal distribution was performed with the software Sigma Stat 3.1 software (Systat Inc, Chicago, IL, USA). For the comparison of the biopsies using surgical templates and the freehand biopsies, the Welch’s t test (two-sample unpooled t test for unequal variances) was performed with the QuickCalcs software (GraphPad Software, Inc. La Jolla, CA, USA).

Results

The mean deviation of the biopsy axes (as described in the materials and methods section) turned out to be 1.3 ± 0.6 mm for the biopsies with a surgical guide and 3.9 ± 1.1 mm for the freehand biopsies (Fig. 9). The data were normally distributed and statistical comparison of the groups revealed a significant difference (p value = 0.0002).

The average biopsy angle deviation was 7.7 ± 4.8° for template biopsies and 16.5 ± 3.8° for freehand biopsies (Fig. 10). The difference of the normally distributed data was again statistically significant (p value = 0.0013).

Fig. 10
figure 10

Box plot of the maximum deviation of axes for the guided and the freehand group. The boxes show the range from the first to the third quartile (*p = 0.0013)

The difference between the planned depth of biopsy channels and the actual depth of biopsy channels was 0.1 ± 2.5 mm for biopsies with surgical guides and 0.4 ± 3.5 mm for freehand biopsies (Fig. 11). These data showed a normal distribution again, but no statistically significant difference was found (p value = 0.85). The time required to fabricate the surgical guides was 128 ± 17 min.

Fig. 11
figure 11

Box plot of the biopsy depth for the guided and the freehand group. The boxes show the range from the first to the third quartile (*p = 0.85)

The data are summarized in the following table (Table 1).

Table 1 Deviation values for biopsies with surgical guides and freehand biopsies

Discussion

This study evaluated the accuracy of in-house 3D-printed surgical guides for biopsies of the lower jaw. Customized 3D-printed models of the lower jaws were used.

The maximum distance of axes turned out to be 1.3 ± 0.6 mm in the guided group and was significantly lower than in the freehand group (3.9 ± 1.1 mm). The angle between the planning and the actual biopsy was significantly lower in the computer-aided group (7.7 ± 4.8°) compared to the freehand group (16.5 ± 3.8°). A significant difference in the accuracy of biopsy depth could not be found when comparing the study group to the control group. No complications or problems occurred during the biopsy procedure in both the study and control group. Due to the lack of data in the literature concerning biopsies with surgical templates, the results of this study can most likely be compared to studies that have examined the accuracy of implant placing with surgical guides printed by stereolithography. Lee et al. report a mean deviation of implants of < 1 mm and mean angular deviation < 3°, however, they measured the distance at the coronal (top) side of the implants [10]. This method of measuring makes sense for implants, but for biopsies the deviation at the end of the biopsy channel is of greater interest. Cassetta et al. found apical deviations of 2.2 ± 0.8 mm and angulation deviations of 4.7 ± 2.7° in their implant study [14]. Van de Wiele et al. reported apical deviations of 1.60 ± 1.7 mm and a difference in angulation of 2.71 ± 2.7° [6]. These deviations are within the range of the present study’s results. The deviation of angles seems higher in our biopsy study compared to evaluations on implants. We suspect that the reason for the increased deviation could be the usually greater distance between the drill channel and the teeth, since the templates in the studies were tooth-supported. The use of templates did not significantly improve the accuracy of biopsy channel depth. This shows that even visual reading of the scale on the trephine drills can be sufficiently accurate.

The use of surgical templates for implant positioning is progressively becoming more popular [15,16,17] and the accuracy of these systems has been recently evaluated by several authors [5,6,7, 9, 10, 14, 18,19,20,21,22,23,24]. Among these evaluations of implant positioning, some of the recent studies also used surgical guides that were printed with a stereolithographic 3D printer [6, 9, 10, 14]. Since the medical technology industry provides these promising systems for dental implants, it is questionable whether these systems are also suitable for an off-label use to perform biopsies of jaws. Until recently, stereolithographic printers with a reasonable resolution, an adequate build volume and biocompatible resins were either not available or not affordable. The desktop printer used in this study has an axis resolution of 0.025 mm and a build volume of 125 × 125 × 165 mm and is, therefore, suitable for producing surgical guides. The printer is able to process a class 1 biocompatible resin. This easily allows the production of a surgical guide which can be used in patients. In our case, the templates were produced in-house, which represents a time-saving and resource-efficient procedure. It has to be considered that conventional surgical guides produced without CAD-CAM technology by dental laboratories are common tools and widely used in dentistry for positioning implants [23]. For dental implants, the axis of implantation can often be determined or approximated via the surface information of the teeth, e.g., from a plaster model [23]. However, a biopsy most importantly requires the osseous information with the location of the lesion, which is needed in addition to the surface information. For the fusion of the osseous information with the surface information, CAD environment is an important tool [25]. Planning of the surgical guide usually results in a very complex 3D body that can be best produced via CAM technology. Thus, CAD/CAM is a very helpful technology for biopsies, and therefore, we think that it is important to evaluate the feasibility and accuracy of this technology in the field of oral surgery.

Dense objects can cause metal artifacts, and therefore, extensive metal dental restorations can complicate planning, which is a limitation of the method [26, 27]. Using 3D-printed lower jaws instead of patients surely presents another limitation of this study. Without soft tissue, the dental and bone geometries were clearly visible and accessible. In contrast, the present control group also has advantages. The dental and osseous geometry of the 3D-printed lower jaws was identical to the lower jaws of real patients with lesions that needed biopsies. This method allowed us to use the same planning for the study group and the control group. This way the potential bias due to different locations and directions of the biopsies and different geometries of the jaws was reduced. Since the freehand osteotomies were performed after the guided biopsies, the surgeon may have already gained an advantage performing the freehand biopsies through practice with the guided biopsies. It can, therefore, be assumed that the accuracy of the freehand biopsies would have been even lower if they had been performed in a completely independent control group. This implies that the statistical significance between guided and freehand biopsies is most likely even higher than we evaluated in this study. The errors of the single steps (e.g., scanning, fusion, printing the guides) were not evaluated. However, this study should evaluate the final result.

Conclusions

From this study, we conclude that surgical guides that were produced with a stereolithographic desktop 3D printer allow significantly higher accuracy of biopsies. Since there were no data in the literature on the accuracy of templates for biopsies in the jaw region, this study is of high relevance. However, the method must now also be investigated in a clinical case series.

Availability of data and materials

Not applicable.

Abbreviations

CAD:

Computer-aided design

CAM:

Computer-aided manufacturing

CT:

Computed tomography

DICOM:

Digital imaging and communications in medicine (file format)

ISO:

International organization for standardization

STL:

Standard tessellation language (file format)

References

  1. Owosho AA, Xu B, Kadempour A, Yom SK, Randazzo J, Ghossein RA, et al. Metastatic solid tumors to the jaw and oral soft tissue: a retrospective clinical analysis of 44 patients from a single institution. J Cranio-Maxillofac Surg. 2016;44(8):1047–53.

    Article  Google Scholar 

  2. Chuong R, Kaban LB. Diagnosis and treatment of jaw tumors in children. J Oral Maxillofac Surg. 1985;43(5):323–32.

    Article  CAS  Google Scholar 

  3. Gaggl A, Schultes G, Kärcher H. Navigational precision of drilling tools preventing damage to the mandibular canal. J Cranio-Maxillofac Surg. 2001;29(5):271–5.

    Article  CAS  Google Scholar 

  4. Valdec S, Schiefersteiner M, Rücker M, Stadlinger B. Guided biopsy of osseous pathologies in the jaw bone using a 3D-printed, tooth-supported drilling template. Int J Oral Maxillofac Surg. 2019;48(8):1028–31.

    Article  Google Scholar 

  5. Naziri E, Schramm A, Wilde F. Accuracy of computer-assisted implant placement with insertion templates. GMS Interdiscip Plastic Reconstr Surg DGPW. 2016;5:doc15.

    Google Scholar 

  6. Van de Wiele G, Teughels W, Vercruyssen M, Coucke W, Temmerman A, Quirynen M. The accuracy of guided surgery via mucosa-supported stereolithographic surgical templates in the hands of surgeons with little experience. Clin Oral Implant Res. 2015;26(12):1489–94.

    Article  Google Scholar 

  7. Geng W, Liu C, Su Y, Li J, Zhou Y. Accuracy of different types of computer-aided design/computer-aided manufacturing surgical guides for dental implant placement. Intern J Clin Exp Med. 2015;8(6):8442–9.

    Google Scholar 

  8. Gao Y, Wu W-J, Zheng L, Huang M-W, Liu S-M, Zhang J-G. Diagnostic value of navigation-guided core needle biopsy in deep regions of the head and neck with focal FDG uptake on 18F-FDG PET/CT. J Cranio-Maxillofac Surg. 2020;48(5):508–13.

    Article  Google Scholar 

  9. Kang SH, Lee JW, Lim SH, Kim YH, Kim MK. Verification of the usability of a navigation method in dental implant surgery: in vitro comparison with the stereolithographic surgical guide template method. J Cranio-Maxillo-Fac Surg Off Publ Eur Assoc Cranio-Maxillo-Fac Surg. 2014;42(7):1530–5.

    Google Scholar 

  10. Lee DH, An SY. Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study. J Adv Prosthodont. 2016;8(3):207–13.

    Article  Google Scholar 

  11. ISO. Geometrical tolerancing—tolerances of form, orientation, location and run-out. ISO 1101: 2004. International Standard Organization, Geneva2005.

  12. Cartiaux O, Paul L, Docquier PL, Francq BG, Raucent B, Dombre E, et al. Accuracy in planar cutting of bones: an ISO-based evaluation. Intern J Med Robot Comput Assist Surg MRCAS. 2009;5(1):77–84.

    Article  Google Scholar 

  13. Postl LK, Kirchhoff C, Toepfer A, Kirchhoff S, Schmitt-Sody M, von Eisenhart-Rothe R, et al. Potential accuracy of navigated K-wire guided supra-acetabular osteotomies in orthopedic surgery: a CT fluoroscopy cadaver study. Intern J Med Robot Comput Assist Surg MRCAS. 2017;13(2):e1752.

    Article  Google Scholar 

  14. Cassetta M, Giansanti M, Di Mambro A, Stefanelli LV. Accuracy of positioning of implants inserted using a mucosa-supported stereolithographic surgical guide in the edentulous maxilla and mandible. Int J Oral Maxillofac Implants. 2014;29(5):1071–8.

    Article  Google Scholar 

  15. D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H, Tahmaseb A. Current state of the art of computer-guided implant surgery. Periodontol 2000. 2017;73(1):121–33.

    Article  Google Scholar 

  16. Dev G, Ferraris F, Arcuri L, Guzzo F, Spinelli D. A novel workflow for computer guided implant surgery matching digital dental casts and CBCT scan. Oral Implantol. 2016;9(1):33–48.

    Google Scholar 

  17. Balasundaram A, Gurun D, Neely A, Ash-Rafzadeh A, Ravichandra J. Novel CBCT and optical scanner-based implant treatment planning using a stereolithographic surgical guide: a multipronged diagnostic approach. Implant Dent. 2014;23(4):401–6.

    Article  Google Scholar 

  18. Lee WC, Huang CH, Chung SC, Wei CC. An efficient and accurate approach for fabricating dental implant surgical guides. Bio-Med Mater Eng. 2014;24(6):2689–95.

    Article  Google Scholar 

  19. Gillot L, Cannas B, Friberg B, Vrielinck L, Rohner D, Pettersson A. Accuracy of virtually planned and conventionally placed implants in edentulous cadaver maxillae and mandibles: a preliminary report. J Prosthet Dent. 2014;112(4):798–804.

    Article  Google Scholar 

  20. Scherer U, Stoetzer M, Ruecker M, Gellrich NC, von See C. Template-guided vs. non-guided drilling in site preparation of dental implants. Clin Oral Investig. 2015;19(6):1339–46.

    Article  Google Scholar 

  21. Dings JPJ, Verhamme L, Maal TJJ, Merkx MAW, Meijer GJ. Reliability and accuracy of skin-supported surgical templates for computer-planned craniofacial implant placement, a comparison between surgical templates: With and without bony fixation. J Cranio-Maxillofac Surg. 2019;47(6):977–83.

    Article  CAS  Google Scholar 

  22. Sun Y, Ding Q, Tang L, Zhang L, Sun Y, Xie Q. Accuracy of a chairside fused deposition modeling 3D-printed single-tooth surgical template for implant placement: an in vitro comparison with a light cured template. J Cranio-Maxillofac Surg. 2019;47(8):1216–21.

    Article  Google Scholar 

  23. Matta R-E, Bergauer B, Adler W, Wichmann M, Nickenig H-J. The impact of the fabrication method on the three-dimensional accuracy of an implant surgery template. J Cranio-Maxillofac Surg. 2017;45(6):804–8.

    Article  Google Scholar 

  24. Shen P, Zhao J, Fan L, Qiu H, Xu W, Wang Y, et al. Accuracy evaluation of computer-designed surgical guide template in oral implantology. J Cranio-Maxillofac Surg. 2015;43(10):2189–94.

    Article  Google Scholar 

  25. Nkenke E, Vairaktaris E, Neukam FW, Schlegel A, Stamminger M. State of the art of fusion of computed tomography data and optical 3D images. Int J Comput Dent. 2007;10(1):11–24.

    CAS  PubMed  Google Scholar 

  26. Bissinger O, Probst FA, Wolff KD, Jeschke A, Weitz J, Deppe H, et al. Comparative 3D micro-CT and 2D histomorphometry analysis of dental implant osseointegration in the maxilla of minipigs. J Clin Periodontol. 2017;44(4):418–27.

    Article  Google Scholar 

  27. Bissinger O, Kirschke JS, Probst FA, Stauber M, Wolff KD, Haller B, et al. Micro-CT vs. whole body multirow detector CT for analysing bone regeneration in an animal model. PLoS ONE. 2016;11(11):e0166540.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors, but was supported by Johannes Kepler Open Access Publishing Fund.

Author information

Authors and Affiliations

Authors

Contributions

LP performed the evaluation and statistical analysis and wrote the manuscript. SH performed the interventions. LP and SH and SK designed the study. All the authors were involved in the data collection, the evaluation and revised the manuscript. All the authors read and approved the final manuscript.

Corresponding author

Correspondence to Lukas Postl.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained from the local ethics committee for this study.

Consent for publication

All the authors gave consent for the publication.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Postl, L., Mücke, T., Hunger, S. et al. In-house 3D-printed surgical guides for osseous lesions of the lower jaw: an experimental study. Eur J Med Res 26, 25 (2021). https://doi.org/10.1186/s40001-021-00495-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40001-021-00495-w

Keywords