Breakage of intramedullary femoral nailing or femoral plating: how to prevent implant failure

Abstract

Introduction

Intramedullary (IM) fixation is the dominant treatment for pertrochanteric and femoral shaft fractures. In comparison to plate osteosynthesis (PO), IM fixation offers greater biomechanical stability and reduced non-union rates. Due to the minimally invasive nature, IM fixations are less prone to approach-associated complications, such as soft-tissue damage, bleeding or postoperative infection, but they are more prone to fat embolism. A rare but serious complication, however, is implant failure. Thus, the aim of this study was to identify possible risk factors for intramedullary fixation (IMF) and plate osteosynthesis (PO) failure.

Materials and methods

We searched our trauma surgery database for implant failure, intramedullary and plate osteosynthesis, after proximal—pertrochanteric, subtrochanteric—or femoral shaft fractures between 2011 and 2019. Implant failures in both the IMF and PO groups were included. Demographic data, fracture type, quality of reduction, duration between initial implantation and nail or plate failure, the use of cerclages, intraoperative microbiological samples, sonication, and, if available, histology were collected.

Results

A total of 24 femoral implant failures were identified: 11 IMFs and 13 POs. The average age of patients in the IM group was 68.2 ± 13.5 years and in the PO group was 65.6 ± 15.0 years, with men being affected in 63.6% and 39.5% of cases, respectively. A proximal femoral nail (PFN) anti-rotation was used in 7 patients, a PFN in one and a gamma nail in two patients. A total of 6 patients required cerclage wires for additional stability. A combined plate and intramedullary fixation was chosen in one patient. Initially, all intramedullary nails were statically locked. Failures were observed 34.1 weeks after the initial surgery on average. Risk factors for implant failure included the application of cerclage wires at the level of the fracture (n = 5, 21%), infection (n = 2, 8%), and the use of an additional sliding screw alongside the femoral neck screw (n = 3, 13%). In all patients, non-union was diagnosed radiographically and clinically after 6 months (n = 24, 100%). In the event of PO failure, the placement of screws within all screw holes, and interprosthetic fixation were recognised as the major causes of failure.

Conclusion

Intramedullary or plate osteosynthesis remain safe and reliable procedures in the treatment of proximal femoral fractures (pertrochanteric, subtrochanteric and femoral shaft fractures). Nevertheless, the surgeon needs to be aware of several implant-related limitations causing implant breakage. These may include the application of tension band wiring which can lead to a too rigid fixation, or placement of cerclage wires at the fracture site.

Introduction

While annual hip fracture incidence declined from 600/100,000 to 400/100,000 person-years from 1996 to 2006, the rarer femur fractures did not decline with incidence rates for subtrochanteric and femoral shaft fractures each below 20 per 100,000 person-years [1]. This incidence of femoral shaft fractures (18.2 per 100,000 person-years [2, 3]), and subtrochanteric fractures (ranging from 10.8 to below 20 per 100,000 person-years) shows a combined incidence below 30 per 100,000 person-years, so this type of fracture is much less common than proximal femur (hip) fracture [4, 5]. Treatment options include plate or intramedullary fixation. AO (Arbeitsgemeinschaft für Osteosynthesefragen, Switzerland) principles emphasise a biologically friendly fixation strategy, anatomical reduction, and adequate micromotion and stiffness at the fracture site to promote healing [6]. The restoration of length, alignment and rotation is emphasised in the fixation of femoral fractures [7].

For treatment, intramedullary nails are less invasive with a lower risk of complications by avoiding damage to the periosteal circulation. In addition, these enable elasticity and, therefore, micro-movement to the facture site. Hereby, the widest and longest nails should be used to achieve a close fit between implant and bones including rotational and lateral stability [8, 9].

Plate osteosynthesis, on the other hand, requires a more open approach, either less invasive or open, to allow an anatomical reduction. It is easier to control the fracture and reduce it through direct visualisation; however, the periosteal circulation may be reduced with the plate pressed onto the bone. Plate osteosynthesis is indicated, especially for comminuted fracture, or in the presence of a narrow femoral shaft, and also for periprosthetic fracture. Besides offering a stable fixation, some stress shielding to the bone occurs, which may induce insufficient strain to the fracture site and, therefore, cause an inadequate healing response associated to non-union, eventually leading to implant failure [10,11,12].

Common complications after fixation include bleeding, malalignment and infection. Modern fixation hardware are associated with low implant failure rates [13]. Intramedullary femoral nails can fail at the proximal screw aperture, as seen in revision nail fixations [14]. In contrast to plate osteosynthesis, intramedullary nailing reaches superior load-to-failure as the implant is closer to the central weight-bearing axis to the femur, which reduces the bending stress by up to 30% [15]. Furthermore, a dynamic stabilisation can be performed to improve compression. On the other hand, disadvantages include the unsuitability of intramedullary for periprosthetic fractures [16]. Plate osteosynthesis has the advantage of allowing compression and direct reduction through the plate in femoral shaft fracture [17, 18].

This study aims to analyse the different causes of implant failure after femoral fracture fixation, including demographics, operative techniques, fracture pattern, the use of cerclage wires, and, if available, histology samples.

Methods

A retrospective study was performed at a major level 1 trauma centre including patients between 2011 and 2019 after obtaining internal review board approval. A fellowship-trained orthopaedic trauma surgeon analysed the internal trauma registry for femoral implant failure and reviewed patient charts as well as plain radiographic imaging. Inclusion criteria consisted of patients aged 18 years or older, suffering from either femoral nail or plate implant failure after per-/subtrochanteric or femoral shaft fracture. Implant failure was defined as either intramedullary nail or plate breakage. Exclusion criteria included simple screw failures as well as the dislocation of implant without failure.

Data on demographics, gender, age and comorbidities were collected. Furthermore, we were interested in time to failure after initial implantation, usage of implant, mechanism of injury (e.g., high versus low energy accidents), location of implant failure and the type of screw fixation (e.g., static versus dynamics stabilisation). Postoperative quality of reduction was defined according to the Lowell’s Criteria in anteroposterior plain radiography which included anteroposterior alignment, displacement and angulation of the two fracture segments [19, 20]. In addition, we were interested in the application of cerclage wires, including the distance between cerclage wires if more than one was applied as well as the distance between A. circumflexa femoris and cerclage wire. In all patients, biopsies were obtained during revision surgery including prolonged microbiology cultures and histology.

For statistical analysis, IBM SPSS and Origin ANOVA was used. Mean and standard error of the mean were calculated assuming the normal distribution of data.

Results

In total, 37 patients with implant failures were found. Twelve were excluded due to injuries in other extremities including the tibia, fibular, humerus, distal radius, clavicular, pelvic plates, and dorsal spondylodesis. In addition, one patient presented with an intramedullary tibial nail failure after a high energy fall, leaving only 24 patients for inclusion. Patients presented with plate breakage in 54.2% (n = 13/24), and intramedullary nail failure in 45.8% (n = 11/24). The overall mean age at time of implant failure was 65.7 ± 15.1 years. Gender was equally distributed at 50.0% between males and females.

Nail implant failure

In the intramedullary nail failure group, 63.6% (n = 7/11) were males and the mean age was 68.2 ± 13.5 years. In this cohort, two patients suffered from diabetes, and two were immunocompromised—one from HIV and one from prednisolone therapy. Furthermore, two people were smokers. Five patients in the IM group were tumour patients (45.5%; 5/11%): two underwent surgery for prophylactic tumour stabilisations (n = 2/11), and three patients received nailing as a fracture treatment—two with prostate cancer and one diffuse large B-cell lymphoma.

The femoral nail breakage occurred at a mean of 36.7 ± 20.3 weeks after initial surgery. All patients presented with sudden onset of pain in the hip, without any trauma. The Synthes proximal femur nail (PFN) was applied in 8 cases (72.7%; n = 8/11), Stryker Gamma nail in two cases (1.8%; n = 2/11) and in one case the nail type was not mentioned (Fig. 1). In all cases, an unreamed long femoral nail was applied with a diameter of 10 mm. The minimum inner cortical diameter of the femoral shafts was 16.1 ± 2.7 mm. All nails except one were locked statically (90.9%; n = 10/11). Cerclages were applied in 6 patients. In 4 patients, a single Synthes cerclage wire system was applied, whereas either a double cable wire configuration or a single cable wire was used in the remaining two. In five of these cases, more than one cerclage wire was used. The distance between two cables was 28.9 mm on average (range from 17 to 54 mm). The most proximal cerclage wire was 24.3 mm on average (range from 0 to 52 mm) away from the anatomical location of the A. circumflexa femoris anterior. In this cohort, subtrochanteric fracture (54.5%; n = 6/11) was the most common, followed by two pertrochanteric (18%; n = 2/11) fracture patterns, one mid shaft fracture (9%; n = 1/11) and one intertrochanteric fracture (9%; n = 1/11). In addition, two patients underwent prophylactic tumour stabilisation with a MIREL score of 9—one for metastasis of melanoma and one for breast cancer.

Fig. 1
figure 1

Two cases of intramedullary implant failure (A) cerclage migration as placed at the fracture site (B) at head screw with adequate placement of doubled wire cerclage and distance more than 5 cm

The initial quality of reduction was anatomic in four cases. In 4 cases, a slight displacement of 7 mm and one case of 8 mm lateralisation was observed on the anteroposterior plain radiographic imaging. In two subjects, no postoperative X-rays were found.

The most common location for implant failure is between two cerclages, as seen in two cases. Breakage was seen at the location of the cephalomedullary screw in three subjects. In three cases, failure sites were close to the most proximal cerclage. In tumour patients, failure occurred at the site of the osteolysis and melanoma metastasis. In those cases, where a cerclage fixation was performed, a dislocation in the fracture was seen (80.0%; n = 4/5) and non-union was observed radiographically in all cases.

At the time of implant failure, the mean serum C-reactive protein was 42.5 ± 68.6 mg/dl and a mean serum white blood cell count of 9.6 ± 3.4 G/l was reported. Histology samples were obtained in two cases which showed non-union without any signs of malignancy or infection. For microbiological cultures, samples remained sterile in 72.7% (n = 8/11); in two patients, the growth of either Staphylococcus aureus (1.8%; n = 2/11) or Streptococcus constellatus (0.9%; n = 1/11) was present. In addition, sonication revealed Streptococcus constellatus in one case, and cutibacterium acnes in another.

Plate implant failure

In the plate osteosynthesis group, the mean age was 65.6 ± 15.0 years at the time of hardware failure. Females made up 61.5% (n = 8/13) of this group. The mean time between initial surgery and implant failure was 21.1 ± 18.9 weeks. In most cases, a Synthes LISS LCP femur plate made of surgical steel was used (n = 12/13). In one patient, a DCP plate was used, and breakage was observed (Fig. 2). In total, plate fixation was utilised for two C3 midshaft fractures, three C3 distal shaft fractures, two prophylactic tumour stabilisations with a MIREL score of 9 (NSCLC), and one clear cell renal cell carcinoma. In addition, plate fixation was also selected for the treatment of two Vancouver C fractures, one interprosthetic fracture, one case of third degree (3°) open C3 femoral shaft fracture, and one subtrochanteric fracture. Locking screws were placed in most cases (92.3%, n = 12/13), while the screws were placed in a nonlocking manner in only one patient. The distance between screw placements was 63.8 ± 52.0 mm and the inner cortical diameter was 16.4 ± 3.3 mm on average. For risk factors, five patients were immunosuppressed with prednisolone (60%; n = 3/5), one was exposed to chemotherapy (i.e., mitomycin) therapy, and one sustained 3° open femoral fracture. Furthermore, one patient suffered from both diabetes mellitus and periprosthetic knee joint infection 1 year prior. No smokers were identified in this cohort. All except one patient suffered from low energy trauma.

Fig. 2
figure 2

Two cases after plate fixation and implant failure with screws close to the fracture site

In all cases, open reduction was performed through the sub-vastus approach and the quality of reduction was anatomic in 5 patients (38.5%; n = 5/13), and minor displacement was seen in three patients ranging between 3 and 7 mm on plain X-ray. Reduction was valgus in three cases, one in varus, and one had a loss of length. For intramedullary failures, non-union was identified in every patient. In three subjects, a single-looped cerclage wires system (Synthes) was applied with a mean distance of 33.5 mm between them. The typical location of implant failure was at the site of fracture close to or with a screw placed inside the fracture (53.8%; n = 7/13). In those cases, where cerclage wires were applied, the implant failure occurred at a location close to them in 23.1% (n = 3/13) of cases with cerclage dislocation. In the remaining cohort, plate breakage was observed for the fixation of an interprosthetic fracture between knee and hip prostheses (n = 2/13; 15.4%), or a progressive osteolysis due to clear cell renal cell carcinoma (7.7%, n = 1/13). The major predisposing factor was the placement of screws in almost every screw hole. The mean serum C-reactive protein was 24.4 ± 33.7 mg/dl and the white blood cell count was 8.4 ± 1.8 G/dl for the plate osteosynthesis group. Histology revealed infection in two cases, whereas microbiology cultures showed the growth of Staphylococcus epidermidis in two cases and Staphylococcus aureus in one. Furthermore, sonication revealed the growth of Staphylococcus hominis in one of the three cases.

All demographics are illustrated in Table 1.

Table 1 Demographics and anatomical variations between nail and plate fixation groups

Discussion

This study outlines the risk factors for the failure of intramedullary as well as plate fixation of the femoral fracture. A trauma database was searched, and clinical charts were reviewed.

These risk factors can be classified into level of trauma, technical factors and biological factors. When looking for the level of trauma and accidental loads, this depends upon individual factors including initial weight bearing after surgery, body mass index, the tendency of falling and involuntary muscle spasms, especially in elderly patients. Hereby, excessive forces beyond the ultimate loads of the fixation strength may act on the implant, causing breakage. Therefore, surgeons should choose the implant as well as the type of surgery according to patient demographics and loading expectations. This may include the usage of thicker implants, stiffer materials, double plating or even a combination of nailing and plating.

More importantly, the type of surgery and excessive implants have to be chosen thoroughly. This includes the length of implant, placement of cerclage wires, locking attachment plates or the necessity of lag screws or narrowly placed screws [7, 21] which may cause local stress concentrations [22,23,24]. Biomechanical investigations showed the weakest spot of the nail to be at the lag screw hole. Forces greater than 1800 N can place the implant at a high risk of failure, especially when drilling off target [25]. Additional weak spots include distal shaft screw holes, where the diameter of the nail is reduced. This also requires an accurate screw placement in either the static or dynamic locking hole to minimise risk [26, 27]. Other factors include the diameter of the femoral nail, as well as a bicortical placement of the distal locking screws. For the placement of cerclage wires, this should be placed from the posterior intermuscular septum at the linea aspera, taking care to avoid a superficial and deep femoral artery. For the distal quarter of the femur, the tip of the cerclage wire passer should be close to the posteromedial and posterior cortex of the femur to prevent injury to the femoral arteries and the sciatic nerve [28]. Other vessels at risk are perforating arteries [29]. An animal study showed a significant decrease in blood circulation in the area between cerclages, raising concerns for non-union [30]. If cerclage wires were placed with a 5 cm space between them, at least one perforator artery was compromised in all cases [31, 32]. Besides poor placement, loose-lock stability can be compromised by surface resorption. One study, however, shows that well tightened cerclages demonstrated in-growth due to callus formation [33]. This includes placement in the fracture line, where it may cause malreduction, leading to instability and, therefore, non-union. Findings suggest that cerclages should be placed at least 1 cm from the tip of the single fracture fragment, especially in oblique fractures. Cerclage wires should be avoided in comminuted and butterfly fracture patterns [24, 33, 34]. Mechanical failure can be avoided by placing a 1 mm cerclage in a mechanically stable region, using a double-looped wire cerclage with a simple symmetrical twist between four and eight times [35, 36]. An increase in wire diameter by 50% increased the load to failure by up to 169% in a single cerclage wire, and up to 300% when doubly wired [37]. In those cases, there is concern for pressure-induced bone necrosis [33, 38]. Furthermore, it is important to be careful during insertion to avoid any damage and notch factors have to be considered which may cause stress-risers, peri-implant fracture and implant fatigue [39]. In plate osteosynthesis, modern plates have limited contact with bone, thus reducing contact pressure and minimising effects on bone blood supply. This enabled a significant decrease in contact area along the interface, to prevent pressure-induced necrosis and subsequent porosis [40]. For bone healing, either absolute or relative stabilisation is required to facilitate primary or secondary bone healing, respectively. In secondary healing, micromotion is required, especially in an axial direction [41]. Therefore, it is important to understand the biomechanics and mechano-biology of the fixation type and implant used. This may also refer to the implant design which could be unsuitably flexible or stress-sensitive.

On the other hand, the biological risk factors including co-morbidities have to be diminished, which requires the fracture pattern, subsequent to an anatomical reduction, to ensure vascularisation, avoid intraoperative fractures and minimise the risk of infection by the adoption of the perioperative antibiotics or, if required, local antibiotics. Comminuted factures can be exposed to substantial stress [14]. In specific cases, augmented biological devices should be used, such as the local application of growth factors, cell therapies or scaffolds.

In our cohort, implant failure occurred, especially in comminuted fractures. In principle, these fracture patterns are already predisposed for delayed or non-union, which was noted in the cases. Cerclage wires were required to achieve a sufficient anatomical reduction.

Besides the patients’ individual, technical and biological risk factors, the life expectancy needs to be considered, especially when stabilising osteolysis, including metastasis and tumours. The longer the lifespan (> 1 year), the more rigid the implant has to be since implants are not indicated for too many repetitive loading cycles. Therefore, the nail thickness, material composition or the use of double plating along with the combination of nail and plate osteosynthesis have to be chosen based upon the individual requirements. In a few cases, the implantation of an endoprosthesis such as individual prosthesis or total femur as well as the application of bone cement (PMMA augmentation) could be an alternative [42].

Conclusion

Fracture fixation is complex and requires some understanding of anatomy, physiology, and biomechanics to achieve union and avoid implant failure. Although implant failure is infrequent, it can have devastating consequences. Therefore, surgeons should minimise potential risk factors by choosing the appropriate implant for a specific patient.

Availability of data and materials

All data are published in this manuscript.

References

  1. Nieves JW, Bilezikian JP, Lane JM, Einhorn TA, Wang Y, Steinbuch M, et al. Fragility fractures of the hip and femur: incidence and patient characteristics. Osteoporos Int. 2010;21:399–408.

    Article  CAS  PubMed  Google Scholar 

  2. Feldstein AC, Black D, Perrin N, Rosales AG, Friess D, Boardman D, et al. Incidence and demography of femur fractures with and without atypical features. J Bone Miner Res. 2012;27:977–86.

    Article  PubMed  Google Scholar 

  3. Sambrook P, Cooper C. Osteoporosis. Lancet. 2006;367:2010–8.

    Article  CAS  PubMed  Google Scholar 

  4. Lee YK, Ha YC, Park C, Yoo JJ, Shin CS, Koo KH. Bisphosphonate use and increased incidence of subtrochanteric fracture in South Korea: results from the National Claim Registry. Osteoporos Int. 2013;24:707–11.

    Article  CAS  PubMed  Google Scholar 

  5. Nieves JW, Cosman F. Atypical subtrochanteric and femoral shaft fractures and possible association with bisphosphonates. Curr Osteoporos Rep. 2010;8:34–9.

    Article  PubMed  Google Scholar 

  6. Gerber C, Mast JW, Ganz R. Biological internal fixation of fractures. Arch Orthop Trauma Surg. 1990;109:295–303.

    Article  CAS  PubMed  Google Scholar 

  7. Sonderegger J, Grob KR, Kuster MS. Dynamic plate osteosynthesis for fracture stabilization: how to do it. Orthop Rev (Pavia). 2010;2:e4.

    Article  Google Scholar 

  8. Zanasi R, Rotolo F, Romano P, Galmarini V, Zanasi L. Intramedullary osteosynthesis. 1. Kuntscher nailing in the femur. Ital J Orthop Traumatol. 1990;16:143–57.

    CAS  PubMed  Google Scholar 

  9. O’Sullivan ME, Chao EY, Kelly PJ. The effects of fixation on fracture-healing. J Bone Joint Surg Am. 1989;71:306–10.

    Article  CAS  PubMed  Google Scholar 

  10. Krettek C, Muller M, Miclau T. Evolution of minimally invasive plate osteosynthesis (MIPO) in the femur. Injury. 2001;32(Suppl 3):SC14-23.

    Article  PubMed  Google Scholar 

  11. Seral B, Garca JM, Doblare M, Seral F. Extramedullary versus intramedullary osteosynthesis in trochanteric hip fractures. Three-dimensional finite-element analysis. Hip Int. 2002;12:150–2.

    Article  CAS  PubMed  Google Scholar 

  12. Charles-Harris M, Lacroix D, Proubasta I, Planell JA. Intramedullary nails vs osteosynthesis plates for femoral fracture stabilisation: a finite element analysis. J Appl Biomater Biomech. 2005;3:157–67.

    CAS  PubMed  Google Scholar 

  13. Petrie J, Sassoon A, Haidukewych GJ. When femoral fracture fixation fails: salvage options. Bone Joint J. 2013;95-B:7–10.

    Article  CAS  PubMed  Google Scholar 

  14. Lambers A, Rieger B, Kop A, D’Alessandro P, Yates P. Implant fracture analysis of the TFNA proximal femoral nail. J Bone Joint Surg Am. 2019;101:804–11.

    Article  PubMed  Google Scholar 

  15. Valverde JA, Alonso MG, Porro JG, Rueda D, Larrauri PM, Soler JJ. Use of the Gamma nail in the treatment of fractures of the proximal femur. Clin Orthop Relat Res. 1998;350:56–61.

    Article  Google Scholar 

  16. Lampropoulou-Adamidou K, Tosounidis TH, Kanakaris NK, Ekkernkamp A, Wich M, Giannoudis PV. The outcome of polyax locked plating system for fixation distal femoral non-implant related and periprosthetic fractures. Injury. 2015;46(Suppl 5):S18-24.

    Article  PubMed  Google Scholar 

  17. Ozkan K, Turkmen I, Sahin A, Yildiz Y, Erturk S, Soylemez MS. A biomechanical comparison of proximal femoral nails and locking proximal anatomic femoral plates in femoral fracture fixation: a study on synthetic bones. Indian J Orthop. 2015;49:347–51.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Zhang WQ, Sun J, Liu CY, Zhao HY, Sun YF. Comparing the intramedullary nail and extramedullary fixation in treatment of unstable intertrochanteric fractures. Sci Rep. 2018;8:2321.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Karanicolas PJ, Bhandari M, Walter SD, Heels-Ansdell D, Sanders D, Schemitsch E, et al. Interobserver reliability of classification systems to rate the quality of femoral neck fracture reduction. J Orthop Trauma. 2009;23:408–12.

    Article  PubMed  Google Scholar 

  20. Leighton RK. Fractures of the neck of the femur. In: Bucholz RW, Heckman JD, Court-Brown CM, editors. Fractures in adults. 6th ed. Lippincott, Philadelphia; 2006. p. 1753–88.

  21. Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ. Biomechanics of locked plates and screws. J Orthop Trauma. 2004;18:488–93.

    Article  PubMed  Google Scholar 

  22. Bhat AK, Rao SK, Bhaskaranand K. Mechanical failure in intramedullary interlocking nails. J Orthop Surg (Hong Kong). 2006;14:138–41.

    Article  CAS  Google Scholar 

  23. Zafiropoulos G, Pratt DJ. Fractured Gamma nail. Injury. 1994;25:331–6.

    Article  CAS  PubMed  Google Scholar 

  24. Najibi S, Mark L, Fehnel D. Mechanical failure of the long gamma nail in two proximal femur fractures. Iowa Orthop J. 2010;30:205–10.

    PubMed  PubMed Central  Google Scholar 

  25. von Ruden C, Hungerer S, Augat P, Trapp O, Buhren V, Hierholzer C. Breakage of cephalomedullary nailing in operative treatment of trochanteric and subtrochanteric femoral fractures. Arch Orthop Trauma Surg. 2015;135:179–85.

    Article  Google Scholar 

  26. Wang CJ, Yettram AL, Yao MS, Procter P. Finite element analysis of a gamma nail within a fractured femur. Med Eng Phys. 1998;20:677–83.

    Article  CAS  PubMed  Google Scholar 

  27. Schoffel N, Moller M, Pross M, Reuther F. Intrapelvic dislocation of lag screw in gamma nailing. Unfallchirurg. 2012;115:750–3.

    CAS  PubMed  Google Scholar 

  28. Apivatthakakul T, Siripipattanamongkol P, Oh CW, Sananpanich K, Phornphutkul C. Safe zones and a technical guide for cerclage wiring of the femur: a computed topographic angiogram (CTA) study. Arch Orthop Trauma Surg. 2018;138:43–50.

    Article  PubMed  Google Scholar 

  29. Gao YS, Guo YJ, Yu XG, Chen Y, Chen C, Lu NJ. A novel cerclage wiring technique in intertrochanteric femoral fractures treated by intramedullary nails in young adults. BMC Musculoskelet Disord. 2018;19:359.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Karakoyun O, Sahin E, Erol MF, Kariksiz M, Kucukkaya M. Effect of cable cerclage on regional blood circulation in rabbits: a scintigraphic study. J Orthop Surg (Hong Kong). 2016;24:367–9.

    Article  Google Scholar 

  31. Apivatthakakul T, Phaliphot J, Leuvitoonvechkit S. Percutaneous cerclage wiring, does it disrupt femoral blood supply? A cadaveric injection study. Injury. 2013;44:168–74.

    Article  CAS  PubMed  Google Scholar 

  32. Angelini A, Battiato C. Past and present of the use of cerclage wires in orthopaedics. Eur J Orthop Surg Traumatol. 2015;25:623–35.

    Article  PubMed  Google Scholar 

  33. Perren SM, Fernandez Dell’Oca A, Lenz M, Windolf M. Cerclage, evolution and potential of a Cinderella technology. An overview with reference to periprosthetic fractures. Acta Chir Orthop Traumatol Cech. 2011;78:190–9.

    CAS  PubMed  Google Scholar 

  34. Wozasek GE, Radler C, Vecsei V. Multiple gamma nail failure. Orthopaedics. 2002;25:183–4.

    Article  Google Scholar 

  35. Shaw JA, Daubert HB. Compression capability of cerclage fixation systems. A biomechanical study. Orthopaedics. 1988;11:1169–74.

    Article  CAS  Google Scholar 

  36. Wahnert D, Lenz M, Schlegel U, Perren S, Windolf M. Cerclage handling for improved fracture treatment. A biomechanical study on the twisting procedure. Acta Chir Orthop Traumatol Cech. 2011;78:208–14.

    CAS  PubMed  Google Scholar 

  37. Bostrom MP, Asnis SE, Ernberg JJ, Wright TM, Giddings VL, Berberian WS, et al. Fatigue testing of cerclage stainless steel wire fixation. J Orthop Trauma. 1994;8:422–8.

    Article  CAS  PubMed  Google Scholar 

  38. Lenz M, Perren SM, Richards RG, Muckley T, Hofmann GO, Gueorguiev B, et al. Biomechanical performance of different cable and wire cerclage configurations. Int Orthop. 2013;37:125–30.

    Article  PubMed  Google Scholar 

  39. Zimmerman KW, Klasen HJ. Mechanical failure of intramedullary nails after fracture union. J Bone Joint Surg Br. 1983;65:274–5.

    Article  CAS  PubMed  Google Scholar 

  40. Lenz M, Perren SM, Gueorguiev B, Richards RG, Krause F, Fernandez Dell’Oca A, et al. Underneath the cerclage: an ex vivo study on the cerclage–bone interface mechanics. Arch Orthop Trauma Surg. 2012;132:1467–72.

    Article  PubMed  Google Scholar 

  41. Uhthoff HK, Poitras P, Backman DS. Internal plate fixation of fractures: short history and recent developments? J Orthop Sci. 2006;11:118–26.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Miller BJ, Soni EE, Gibbs CP, Scarborough MT. Intramedullary nails for long bone metastases: why do they fail? Orthopaedics. 2011;34:4.

    Article  Google Scholar 

Download references

Acknowledgements

No.

Funding

Open Access funding enabled and organized by Projekt DEAL. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. No funds were received in support of this study.

Author information

Authors and Affiliations

Authors

Contributions

HCB and KFB designed and directed the project. HCB developed the theory and performed the computations. CW, MH, HCB and KFB verified the analytical methods. US and CP supervised the findings of this work. All authors discussed the results and contributed to the final manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Henrik C. Bäcker.

Ethics declarations

Ethics approval and consent to participate

We obtained internal review board approval. Informed consent was waived because of the retrospective design.

Competing interests

The authors have no competing interest directly related to this work.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bäcker, H.C., Heyland, M., Wu, C.H. et al. Breakage of intramedullary femoral nailing or femoral plating: how to prevent implant failure. Eur J Med Res 27, 7 (2022). https://doi.org/10.1186/s40001-021-00630-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40001-021-00630-7

Keywords