Effect of lung flooding and high-intensity focused ultrasound on lung tumours: an experimental study in an ex vivo human cancer model and simulated in vivo tumours in pigs
European Journal of Medical Research volume 19, Article number: 1 (2014)
Abstract
Background
High-intensity focused ultrasound is a valuable tool for minimally invasive tumour ablation. However, due to the air content in ventilated lungs, lung tumours have never been treated with high-intensity focused ultrasound. Lung flooding enables efficient lung sonography and tumour imaging in ex vivo human and in vivo porcine lung cancer models. The current study evaluates the effectiveness of lung flooding and sonography-guided high-intensity focused ultrasound for lung tumour ablation in ex vivo human and in vivo animal models.
Methods
Lung flooding was performed in four human lung lobes which were resected from non-small cell lung cancers. B-mode imaging and temperature measurements were simultaneously obtained during high-intensity focused ultrasonography of centrally located lung cancers. The tumour was removed immediately following insonation and processed for nicotinamide adenine dinucleotide phosphate-diaphorase and H&E staining. In addition, the left lungs of three pigs were flooded. Purified BSA in glutaraldehyde was injected centrally into the left lower lung lobe to simulate a lung tumour. The ultrasound was focused transthoracically through the flooded lung into the simulated tumour with the guidance of sonography. The temperature of the tumour was simultaneously measured. The vital signs of the animal were monitored during the procedure.
Results
A well-demarcated lesion of coagulation necrosis was produced in four of four human lung tumours. There did not appear to be any damage to the surrounding lung parenchyma. After high-intensity focused ultrasound insonation, the mean temperature increase was 7.5-fold higher in the ex vivo human tumour than in the flooded lung tissue (52.1 K ± 8.77 K versus 7.1 K ± 2.5 K). The transthoracic high-intensity focused ultrasound of simulated tumours in the in vivo model resulted in a mean peak temperature increase up to 53.7°C (±4.5). All of the animals survived the procedure without haemodynamic complications.
Conclusions
High-intensity focused ultrasound with lung flooding produced a thermal effect in an ex vivo human lung carcinoma and in vivo simulated lung tumours in a porcine model. High-intensity focused ultrasound is a potential new strategy for treating lung cancer.
Background
Lung cancer remains the leading cause of cancer-related death in the western population [1, 2]. In addition to primary tumours, the lung is the most common site for metastatic spread of other malignancies. Curative treatment of lung tumours or metastases requires surgical resection, which results in the loss of functional lung parenchyma. Unfortunately, fewer than half of patients are eligible for curative surgery due to limited lung function [3]. Radiation or chemotherapy alone rarely cures patients. Minimally invasive local ablation (for example, percutaneous laser or radiofrequency) results in major complications in 6% of patients and procedure-specific 30-day mortality rates of 2.6% [4]. Furthermore, the local progression rate is 35% [5]. Therefore, these procedures are palliative strategies.
High-intensity focused ultrasound (HIFU) is a non-invasive, highly precise procedure to locally destroy tissue through ablation. Current clinical trials are evaluating the effectiveness of HIFU for the treatment of cancers of the brain, breast, liver, bone, and prostate [6–9]. However, lung cancers have never been treated with this approach, because the ventilated lung is a total acoustic absorber and reflector [10]. This problem was solved by lung flooding. Lung flooding enables efficient lung sonography and tumour imaging in ex vivo human and in vivo porcine lung cancer models [11]. The conditions for applying HIFU to lung tumours are provided.
The aim of this study was to evaluate the effectiveness of sonography-guided HIFU with lung flooding for lung tumour ablation in ex vivo human and in vivo animal models.
Methods
Ex vivo examinations
Preparation of resected human lung lobes with carcinoma
Four patients with non-small cell lung cancer (NSCLC) received therapeutic lobectomies of the right lower lobe (n = 2), right upper lobe (n = 1), or left lower lobe (n = 1). The average age of the patients was 75.7 years (range 68 to 82 years). The mean tumour diameter was determined by computed tomography and was 4.1 cm (range 3.2 to 5.1 cm). Pre-surgical histological diagnoses of three adenocarcinomas and one large cell neuroendocrine carcinoma were confirmed with percutaneous or transbronchial needle biopsies. The lobes were cooled in a 15°C isotonic liquid bath immediately following resection and prepared for liquid filling (flooding) in tabula. An expanded polytetrafluoroethylene graft was then anastomosed end-to-end with the lobar bronchi. An infusion system was connected to the conduit, and the lobes were continuously filled with isotonic saline (15°C) until the functional residual capacity was reached. The lung lobe was examined transpleurally by ultrasound (MicroMaxx™ Portable Ultrasound System; SonoSite, Inc., Bothell, WA, USA) in the liquid bath with a linear probe (L38e, 10 to 5 MHz; SonoSite, Inc., Bothell, WA, USA) in fundamental B-mode. The lobes were selected for HIFU only in cases with gas-free filling and complete sonographic imaging of the tumour. This study was approved by the ethics committee of the Medical Association of Thuringia.
HIFU application
The experimental HIFU setup contained the HIFU transducer H102 (Sonic Concepts Inc., Bothell, WA, USA), a power amplifier (RF Source, Athena, Greece), a Type K thermocouple (XF339, Labfacility Ltd., Sheffield, South Yorkshire, UK), and a digital multimeter (34401A, Agilent, Santa Clara, CA, USA). The HIFU setup and alignment with the tumour are illustrated in Figure 1. The thermocouple was guided through the centre hole of the HIFU transducer into the tumour. The alignment of the HIFU focal zone within the tumour and the thermocouple was monitored by ultrasound with a curved imaging probe (C11e, SonoSite, Inc., Bothell, WA, USA), which was mounted sideways on the HIFU transducer. Self-heating of the thermocouple was excluded by applying HIFU to the thermocouple in water. An impedance-matching network provided by the manufacturer drives the HIFU probe with a radiofrequency (RF) amplifier. The acoustic power of the focal zone in water was estimated to be 2,500 W/cm2 at 1.1 MHz (ISPTP, spatial peak, temporal peak) based on the manufacturer’s calibration sheet. A duty cycle of 50% and a repetition frequency of 10 KHz were applied. The cancer tissue was exposed to HIFU for ten seconds. The focal zone was immediately moved through the tumour in lateral 2 mm increments. HIFU was initiated again for ten seconds, and ultrasound B-mode imaging was then performed. HIFU exposure ended after one slice of the HIFU zone moved through the tumour volume. Subsequently, the HIFU focus and thermocouple were placed into flooded lung tissue. The procedure was repeated as described above. All experimental steps were performed within 45 minutes of surgical resection.
Histopathology and enzyme histochemistry
The specimen, including the tumour, was cut at a 90° angle to the treated cross section immediately following the last HIFU exposure. Each half of the tumour was sectioned parallel to the cutting surface to show the transition from ablated to non-ablated tumour. One part was fixed in 4% formaldehyde for H&E, and the other was frozen in liquid nitrogen for nicotinamide adenine dinucleotide phosphate (NADPH)-diaphorase staining. An intralobar metastasis from one case served as an untreated tissue control.
In vivo examination
Animal preparation
Animal experiments were performed on three female pigs (Deutsches Landschwein breed; weight range: 33 to 38 kg, average: 35.5 kg) with permission from the Veterinary Department of the Thuringian State Authority for Food Protection and Fair Trading in compliance with the National Animal Protection Act. Total intravenous anaesthesia was initiated with propofol (10 mg/kg/h), fentanyl (0.05 to 0.08 μg/kg/min), and pancuronium bromide (2.5 μg/kg/min). A left-sided Robertshaw double-lumen tube with an extra-long bronchial lane (size 39 Ch; special product by Mallinckrodt Medical, Dublin, Ireland) was inserted after tracheotomy. Mechanical ventilation was performed with an ICU respirator (Servo 900, Siemens AG, Munich, Germany) on a volume-controlled setting (FIO2 = 1.0; tidal volume 10 ml/kg; respiratory rate 16 to 20/min; positive end expiratory pressure = 6 cm H2O). The electrocardiogram, arterial blood pressure, capillary oxygen saturation, and expiratory CO2 concentration were measured and recorded continuously (Datex AS/3 Compact Multiparameter Patient Monitor; Datex-Ohmeda Corp., Helsinki, Finland). Arterial blood gases were analysed every 30 minutes (ABL System 625; Radiometer Medical, Copenhagen, Denmark).
Tumour simulation and HIFU application
The HIFU setup for transthoracic application on pigs contained a self-manufactured sample holder. The HIFU transducer (SU102, Sonic Concepts Inc., Bothell, WA, USA) was arranged with the sideways attached puncture channel to guide the thermocouple and curved imaging probe (C11e, MicroMaxx; SonoSite Inc., Bothell, WA, USA). The three elements were arranged such that the thermocouple tip was in the focus of the HIFU beam. A rib was resected after lung flooding without pleural injuries. The HIFU applicator was placed above the resected rib (Figure 2). A 17-G needle was placed in the centre of the lower left lung lobe with the guidance of a percutaneous transpleural ultrasound. Five millilitres of a fluid composed of purified BSA and glutaraldehyde (Bioglue®; CryoLife Europa, Guildford, UK) were injected to simulate a lung tumour. A thermocouple was immediately inserted through the needle into the simulated tumour. The alignment of the focal zone with the target lesion and thermocouple was monitored by ultrasound imaging. The surgical HIFU probe with an outer diameter of 35 mm operates at 3.0 MHz. The focal acoustic intensity was estimated to be 2,000 W/cm2 MHz (ISPTP , spatial peak, temporal peak) based on the manufacturer’s calibration sheet. HIFU was applied according to a ‘one second on, one second off’ scheme. The temperature inside the simulated tumour was recorded by the thermocouple during HIFU exposure. Data were transferred and stored on a laptop.
Results
Ex vivo examinations
The mean temperature in the tumour tissue increased by 52.1 K ± 8.77 K after ten seconds of HIFU exposure. The same acoustic intensity produced a mean temperature increase of 7.1 K ± 2.5 K in a flooded lung. The temperature increase in the ex vivo human NSCLC was 7.5-fold higher than that in the flooded lung tissue (Figure 3). The sonoablated cancer tissue appeared to be strongly hyperechoic immediately following HIFU application (Figure 4). The strength of echogenicity decreased 30 seconds after HIFU exposure but did not disappear. The ablated cancer tissue appeared to be slightly whitish in all four cases. A demarcation between treated and non-treated areas was observed (Figure 5).
H&E staining showed coagulative necrosis with cytoplasmic eosinophilia, disruption of cellular membranes with blurred cytoplasmic borders, karyolysis, and nuclear pyknosis. Insonated and non-insonated cancer tissues showed sharp demarcations. There was a small region between the area of coagulation necrosis and vital cancer tissue in which cells showed cytoplasmic vacuolization (Figure 6). These characteristics were not found in the untreated areas of cancer tissue. NADPH-diaphorase staining produced a sharp demarcation between ablated and non-ablated cancer tissue, and demonstrated that vital lung tissue was present immediately adjacent to ablated cancer tissue (Figure 7). HIFU-treated tumour tissue did not show evidence of vitality by NADPH-diaphorase staining. Sonographic lesions were not detected in the tumour surrounding the lung tissue or in the direction of the HIFU irradiation. In contrast, a non-treated metastatic tumour in the same lobe showed complete vitality.
In vivo examination
A lung tumour could be simulated in the flooded lung of each of the three animals by sonography-guided injection of BioGlue®. The central tumour appeared hypoechoic by B-mode sonography and was sharply demarcated from the lung parenchyma with small inclusions of air. A temperature rise was monitored within each one-second application of HIFU irradiation. A section of 25 seconds of HIFU exposure in the ‘one second on, one second off’ scheme is shown in Figure 8. Repeated HIFU insonation caused an overall average temperature increase of up to 53.7°C (±4.5, n = 3).
Arterial blood pressure, oxygen, and carbon dioxide partial pressure were normal in all three animals during and after HIFU insonation. No animals died during the procedure.
Discussion
Currently, the best curative treatment for early-stage non-small cell lung cancer and lung metastases is surgical resection. However, surgery is not appropriate for patients with reduced pulmonary function due to the loss of lung parenchyma. Minimally invasive technologies for local tumour control have been developed as alternatives to surgery. The most common method is percutaneous radiofrequency ablation. However, this kind of thermal ablation is limited because the extent to which tissue has been successfully ablated cannot be accurately assessed. This could be a reason for the observed rate of local recurrence. Another risk factor for local tumour progression is the adherence of viable tumour tissue to needle applicators and subsequent spreading of tumour cells [12]. In addition, pulmonary percutaneous radiofrequency ablation is invasive and associated with major complications and mortality.
HIFU has been examined in animal experiments and clinical trials as a technique for localized tissue ablation for more than 60 years. HIFU is superior to other radiation beams, because it penetrates deep and selectively destroys tumour tissue. In addition, HIFU can be applied as many times as needed [13]. HIFU is currently used as a therapy for cancers of the abdominal organs, breast, and brain [14–18].
Non-invasive local therapies with curative intention are not available for lung tumours. HIFU has never been applied to lung tumours, because the air content in ventilated lungs reflects acoustic intensities. We solved this problem with lung flooding. Lung flooding enables efficient lung sonography and tumour imaging in ex vivo human and in vivo porcine lung cancer models [11]. The cancerous tissue was visualized by replacing the alveolar gas with fluid, facilitating ultrasound-guided HIFU. The aim of the present study was to explore the ability of therapeutic ultrasound to penetrate the overlying lung parenchyma without damage and to increase the temperature of the tumour tissue.
A tissue temperature greater than 60°C will usually cause instantaneous and irreversible cell death within one second in most extrapulmonary tissues due to coagulation necrosis. Coagulation necrosis is the primary mechanism by which HIFU destroys tumour cells [6]. Our ex vivo studies showed that lung cancer tissue absorbed acoustic energy, leading to an increase in tumour temperature of 52.1 K. Based on the temperature of a cooled lung lobe (15°C), a peak temperature of 67°C was created. The heat induced by HIFU was sufficient to generate coagulation necrosis in the lung cancer tissue. In contrast, the same acoustic HIFU energy caused a minimal temperature increase (7.1 K) in flooded parenchyma. NADPH-diaphorase staining showed that HIFU was highly selective for tumour tissue in flooded lungs. The lung parenchyma directly adjacent to the ablated cancer tissue was undamaged and viable. This can be explained by the very low attenuation of saline, which is the major mass component in flooded lungs [19]. Thus, the risk of damaging the surrounding tissue or adjacent organs can be minimised as previously described in a study of HIFU in abdominal tumours [20]. A normal saline solution was injected into the abdominal cavity to reduce complications from HIFU therapy for abdominal tumours. This method does not have adverse effects on the efficacy of HIFU ablation [21].
In vivo transthoracic HIFU application in pigs caused a mean peak temperature increase up to 53.7°C in a simulated lesion (BioGlue®) deep inside the flooded lung. HIFU energy penetrated through the pleura and flooded lung into the target lesion. The simulated lesion, which was located at a 6 cm depth below the transducer, was heated by HIFU. The temperature increase was highly variable and inconstant. The mechanism of heat generation in the BioGlue®-simulated lesion was unclear. The lesion consisted of purified BSA and glutaraldehyde. Although this simulated lesion was not a true human tumour, the acoustic and thermal properties are similar to those of human cancerous tissue due to similar densities and high protein contents. In addition, movements of the heart and mediastinum may be transmitted to the target lesion, bringing them out of the focal zone.
Lung flooding seems to be ideal for HIFU application, because sonographic imaging is possible and because of other advantages.
Compared to other human tissues, a flooded lung has an ideally suitable beam path, because water has a very low attenuation. In contrast to flooded lung tissue, tumour tissue converts acoustic energy into a therapeutic thermal dose and enables selective heating of the cancer mass. Consequently, damage to healthy lung tissue is minimised.
In addition to acoustic advantages, there are other favourable conditions for tumour ablation in a flooded lung. There is no pulmonary blood flow in a flooded lung [22]. Perfusion reduces heat and is not desirable with locoregional thermal therapy. In addition, ischaemia-related acidosis sensitises tumour tissue to heat. In comparison to a ventilated lung, lung flooding reduces tumour movements caused by breathing.
The following limitations existed in the current study.
The risk of lung flooding in patients with limited lung function is unclear. Further studies on patients with limited lung function should be performed to examine the influence of unilateral lung flooding on haemodynamics and gas exchange. Ongoing studies in pigs showed that lung flooding of only one lobe is feasible.
Transcutaneous application was not applied for the in vivo or ex vivo studies. One rib had to be resected to apply the HIFU applicator, which consisted of a HIFU transducer and sector-array probe, to the chest of a pig. Technical improvements in the application system are necessary to ensure that the HIFU focal zone is aimed precisely at the target lesion with the guidance of sonography. Transthoracic HIFU application is difficult despite resection of the ribs, because the intercostal spaces in the animal model are narrow. A method and transducers have been developed to avoid the shielding of therapeutic ultrasound by the human rib cage [23].
Different HIFU exposure schemas (ten seconds ex vivo versus ‘one second on/off’ in vivo) and transducers were applied during this study. Ultrasound imaging is disturbed if HIFU is continuously on. Therefore, the intermittent ‘one second on/off’ in vivo schema was needed to control the focal alignment with the thermocouple during the ‘one second off’ interval.
For HIFU application to human tumours, resected lung lobes were used.
The gas-free filling occurs at 71.4% of the resected lobes. This limitation is due to residual air in non-collapsed bronchi that can only solved by resorption under in vivo conditions [11]. Lung flooding was performed with cooled saline at 15°C for the ex vivo HIFU application to ensure that hypoxic damage did not occur within the tumour and lung tissues. This was important, because the NADPH-diaphorase staining method is based on mitochondrial vitality, which is very sensitive to ischaemia.
A hyperechoic area was found within the tumour tissue immediately following HIFU exposure. Grayscale changes have been shown to be a useful marker for HIFU-induced tissue destruction [8]. However, it is unclear whether the area of grayscale changes corresponds with the ablated area, such that a hyperechoic sonolesion represents irreversibly damaged tissue. B-mode ultrasound imaging is probably not the best method for monitoring tissue response. Currently, the most important problem associated with ultrasound-guided HIFU ablation is the lack of reliable thermometry and lesion production monitoring [6]. In addition to ultrasound imaging, real-time magnetic resonance thermometry could be important for ascertaining the extent of tumour destruction [24–26].
Conclusions
In combination with lung flooding, high-intensity focused ultrasound produced a thermal effect in an ex vivo model of human lung carcinoma and in simulated lung tumours in an in vivo porcine model. High-intensity focused ultrasound is a potential strategy for treating lung cancer. Further studies will examine HIFU therapy in animal models of lung cancer.
Abbreviations
- BSA:
-
Bovine serum albumin
- FIO2:
-
Fraction of inspired oxygen
- HIFU:
-
High-intensity focused ultrasound
- H&E:
-
Hematoxylin and eosin
- NSCLC:
-
Non-small cell lung cancer
- NADPH-diaphorase:
-
Nicotinamide adenine dinucleotide phosphate-diaphorase.
References
DKFZ: Atlas of Cancer Mortality. Heidelberg: German Center of Cancer Research; 2007.
Drings P, Dienemann H, Wannenmacher M: Management of Lung cancer. Berlin, Heidelberg, New York: Springer; 2002.
Colice GL, Rubins J, Unger M: Follow-up and surveillance of the lung cancer patient following curative-intent therapy. Chest 2003, 123: 272–283. 10.1378/chest.123.1.272
Simon CJ, Dupuy DE, DiPetrillo TA, Safran HP, Grieco CA, Ng T, Mayo-Smith WW: Pulmonary radiofrequency ablation: long-term safety and efficacy in 153 patients. Radiology 2007, 243: 268–275. 10.1148/radiol.2431060088
Von Meyerfeldt EM, Prevoo W, Peyrot D, Lai a Fat N, Burgers SJ, Wouters MW, Klomp HM: Local progression after radiofrequency ablation of pulmonary metastases. Cancer 2011, 117: 3781–3787. 10.1002/cncr.25958
Zhou YF: High intensity focused ultrasound in clinical tumour ablation. World J Clin Oncol 2011, 2: 8–27. 10.5306/wjco.v2.i1.8
Crouzet S, Rebillard X, Chevallier D, Rischmann P, Pasticier G, Garcia G, Rouviere O, Chapelon JY, Gelet A: Multicentric oncologic outcomes of high-intensity focused ultrasound for localized prostate cancer in 803 patients. Eur Urol 2010, 58: 559–566. 10.1016/j.eururo.2010.06.037
Nerendra T, Sanghvi T, Rescorla F, Galliani C, Fry F, Griffith S, Grosfeld J, Yang R: Extracorporeal liver ablation using sonography-guided high intensity focused ultrasound. Investig Radiol 1992, 27: 796–803. 10.1097/00004424-199210000-00009
Crouzet S, Poissonnier L, Murat FJ, Pasticier G, Rouvière O, Mège-Lechevallier F, Chapelon JY, Martin X, Gelet A, Melodelima D, Vincenot J, Chen Y, Dupre A, Chapelon J, Rivoire M: Washington: outcomes of HIFU for localised prostate cancer using the ablatherm integrate imaging® device. Prog Urol 2011, 21: 191–197. 10.1016/j.purol.2010.07.005
Mikhak Z, Pedersen P: Acoustic attenuation properties of the lung: an open question. Ultrasound Med Biol 2002, 28: 1209–1216. 10.1016/S0301-5629(02)00561-6
Lesser TG, Schubert H, Bischoff S, Wolfram F: Lung flooding enables efficient lung sonography and tumour imaging in human ex vivo and porcine in vivo lung cancer model. Eur J Med Res 2013, 18: 23. [Epub ahead of print] 10.1186/2047-783X-18-23
Snoeren N, Huiskens J, Rijken AM, van Hillegersberg R, van Erkel AR, Slooter GD, Klaase JM, van den Tol GM, Ten Kate F, Jansen MCJ, van Gulik TM: Viable tumour tissue adherent to needle applicators after local ablation: a risk factor for local tumour progression. Ann Surg Oncol 2011, 18: 3702–3710. 10.1245/s10434-011-1762-8
Kennedy J, ter Haar G, Cranston D: High intensity focused ultrasound: surgery of the future? Br J Radiol 2003, 76: 590–599. 10.1259/bjr/17150274
Thüroff S, Chaussy C, Vallancien G, Wieland W, Kiel HJ, Le Duc A, Desgrandchamps F, De La Rosette JJ, Gelet A: High-intensity focused ultrasound and localized prostate cancer: efficacy results from the European multicentric study. J Endourol 2003, 17: 673–677. 10.1089/089277903322518699
Furusawa H, Namba K, Nakahara H, Tanaka C, Yasuda Y, Hirabara E, Imahariyama M, Komaki K: The evolving non-surgical ablation of breast cancer: MR guided focused ultrasound (MRgFUS). Breast Cancer 2007, 14: 55–58. 10.2325/jbcs.14.55
Wu F, Wang ZB, Zhu H, Chen WZ, Zou JZ, Bai J, Li KQ, Jin CB, Xie FL, Su HB: Feasibility of US-guided high-intensity focused ultrasound treatment in patients with advanced pancreatic cancer: initial experience. Radiology 2005, 236: 1034–1040. 10.1148/radiol.2362041105
Wu F, Wang ZB, Chen WZ, Zou JZ, Bai J, Zhu H, Li KQ, Jin CB, Xie FL, Su HB: Advanced hepatocellular carcinoma: treatment with high-intensity focused ultrasound ablation combined with transcatheter arterial embolization. Radiology 2005, 235: 659–667. 10.1148/radiol.2352030916
Chen W, Zhou K: High-intensity focused ultrasound ablation: a new strategy to manage primary bone tumours. Curr Opin Orthop 2005, 16: 494–500. 10.1097/01.bco.0000183682.00083.a1
Treeby B, Zhang E, Thomas AS, Cox B: Measurement of the ultrasound attenuation and dispersion in whole human blood and its components from 0 to 70 MHz. Ultrasound Med Biol 2011, 37: 289–300. 10.1016/j.ultrasmedbio.2010.10.020
Yang R, Reilly CR, Rescorla FJ, Faught PR, Sanghvi NT, Fry FJ, Franklin TD Jr, Lumeng L, Grosfeld JL: High-intensity focused ultrasound in the treatment of experimental liver cancer. Arch Surg 1991, 126: 1002–1009. 10.1001/archsurg.1991.01410320088012
Wu CC, Chen WS, Ho MC, Huang KW, Chen CN, Yen JY, Lee PH: Minimizing abdominal wall damage during high intensity focused ultrasound ablation by inducing artificial ascites. J Acoust Soc Am 2008, 124: 674–679. 10.1121/1.2839907
Lesser T: Neue Möglichkeiten der kompletten Lungensonographie bei einseitiger Flüssigkeitsfüllung: experimentelle Untersuchungen zur Technik, Herdsimulation, Hämodynamik und Morphologie. Jena: Universitätsbibliothek; 1999.
Rieke V, King RL, Swaminathan A, McConnell M, Butts Pauly K: MR-guided focused ultrasound ablation through the ribcage. Intl Soc Mag Reson Med 2009, 17: 2529.
Jolesz FA, Hynynen K, McDannold N, Tempany C: MR imaging-controlled focused ultrasound ablation: a noninvasive image-guided surgery. Magn Reson Imaging Clin N Am 2005, 13: 545–560. 10.1016/j.mric.2005.04.008
Ellis S, Rieke V, Kohi M, Westphalen AC: Clinical applications for magnetic resonance guided high intensity focused ultrasound (MRgHIFU): present and future. J Med Imaging Radiat Oncol 2013, 57: 391–399. 10.1111/1754-9485.12085
Rieke V, Butts Pauly K: MR thermometry. J Magn Reson Imaging 2008, 27: 376–390. 10.1002/jmri.21265
Acknowledgments
This work was funded by the SRH Foundation, Heidelberg, Germany. The funder had no role in the study design or in the collection, analysis, and interpretation of data. The authors would also like to thank Uwe Leder for his organizational support and Mrs. Petra Dobermann for active help with the animal experiments.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no financial or non-financial competing interests.
Authors’ contributions
FW was responsible for HIFU technique and temperature measurement. He collected and evaluated the data and wrote the manuscript. HS and SB performed the anaesthesia. CB performed the histopathological and enzyme histochemistry examinations. TGL co-wrote and revised the manuscript and discussed the results with the authors. All authors read and approved the manuscript.
Authors’ original submitted files for images
Below are the links to the authors’ original submitted files for images.
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Wolfram, F., Boltze, C., Schubert, H. et al. Effect of lung flooding and high-intensity focused ultrasound on lung tumours: an experimental study in an ex vivo human cancer model and simulated in vivo tumours in pigs. Eur J Med Res 19, 1 (2014). https://doi.org/10.1186/2047-783X-19-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/2047-783X-19-1