Rapid point-of-care detection of SARS-CoV-2 infection in exhaled breath using ion mobility spectrometry: a pilot study

Abstract

Background

An effective testing strategy is essential for pandemic control of the novel Coronavirus disease 2019 (COVID-19) caused by infection with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Breath gas analysis can expand the available toolbox for diagnostic tests by using a rapid, cost-beneficial, high-throughput point-of-care test. We conducted a bi-center clinical pilot study in Germany to evaluate breath gas analysis using multi-capillary column ion mobility spectrometry (MCC-IMS) to detect SARS-CoV-2 infection.

Methods

Between September 23, 2020, and June 11, 2021, breath gas measurements were performed on 380 patients (SARS-CoV-2 real-time polymerase chain reaction (PCR) positive: 186; PCR negative: 194) presenting to the emergency department (ED) with respiratory symptoms.

Results

Breath gas analysis using MCC-IMS identified 110 peaks; 54 showed statistically significant differences in peak intensity between the SARS-CoV-2 PCR-negative and PCR-positive groups. A decision tree analysis classification resulted in a sensitivity of 83% and specificity of 86%, but limited robustness to dataset changes. Modest values for the sensitivity (74%) and specificity (52%) were obtained using linear discriminant analysis. A systematic search for peaks led to a sensitivity of 77% and specificity of 67%; however, validation by transferability to other data is questionable.

Conclusions

Despite identifying several peaks by MCC-IMS with significant differences in peak intensity between PCR-negative and PCR-positive samples, finding a classification system that allows reliable differentiation between the two groups proved to be difficult. However, with some modifications to the setup, breath gas analysis using MCC-IMS may be a useful diagnostic toolbox for SARS-CoV-2 infection.

Trial registration: This study was registered at ClinicalTrials.gov on September 21, 2020 (NCT04556318; Study-ID: HC-N-H-2004).

Background

Almost 3 years after its emergence, coronavirus disease 2019 (COVID-19), caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), remains a global burden [1]. In addition to developing vaccines and therapeutic options, an effective testing strategy remains an important component of the pandemic response, particularly with new variants constantly emerging [2, 3]. Fast and reliable test results are essential for pandemic control, especially in healthcare facilities. Particularly in the Emergency Department (ED) setting, screening strategies are key for intra-hospital patient flow, isolation precautions, and avoiding nosocomial outbreaks [4, 5]. To follow antimicrobial stewardship (AMS) concepts, rapidly identifying a viral pathogen as the source of an acute infection is paramount to avoiding antibiotic overuse [6, 7].

Real-time polymerase chain reaction (RT-PCR) from nasopharyngeal swabs is the gold standard for SARS-CoV-2 diagnosis [8, 9]. However, laboratory-based RT-PCR with a turnaround time (TAT) of more than four hours is not applicable for rapid decision-making in the ED, and upcoming point-of-care RT-PCRs with TATs of approximately 20 min minimum are too costly and limited in their availability [10]. SARS-CoV-2 antigen testing can be performed at the point of care and usually provides results after approximately 15 min. However, it should be noted that the sensitivity of antigen testing is significantly lower compared to RT-PCR, especially at low viral loads [11,12,13]. Moreover, new SARS-CoV-2 variants can escape detection by commonly used antigen tests [14, 15]. Microfluidics-based testing represents a promising future technology, but application at the point of care is still in the early stages of development [16, 17]. Thus, there is a great need for rapid and resource-saving diagnostic alternatives that can be performed at point of care.

In SARS-CoV-2 infection, the upper airways are the main sites of viral replication and inflammation. Therefore, it seems reasonable that the analysis of exhaled breath could represent an attractive, non-invasive approach [18]. Breath gas analysis can quickly and reliably detect volatile organic compounds (VOCs) at their lowest concentrations. The most common methods are ion mobility spectrometry (IMS) and mass spectrometry (MS) [19,20,21,22]. In recent years, breath gas analysis has been used to detect various respiratory pathogens [23,24,25,26,27].

The first attempts have already been made to identify a SARS-CoV-2-specific fingerprint in the VOC profile of breath [18]. In a preclinical study, we demonstrated that IMS can discriminate SARS-CoV-2 from other respiratory viruses by analyzing air samples collected from the headspace of virus-infected in vitro cultures [28]. The first small clinical studies on the detection of SARS-CoV-2 in breath gas showed promising results [29,30,31,32,33,34,35,36]. For instance, Ibrahim et al. identified seven exhaled breath features (benzaldehyde, 1-propanol, 3,6-methylundecane, camphene, beta-cubebene, iodobenzene, and an unidentified compound) that distinguished between PCR-positive and PCR-negative patients using regression analysis [32]. In another study by Grassin-Delyle et al. [33] the compounds methylpent-2-enal, 2,4-octadiene, 1-chloroheptane, and nonanal were found to be useful in identifying patients with COVID-19-associated acute respiratory distress syndrome (ARDS).

However, because of the high inter-individual diversity and variability of VOC profiles, studies with a large number of patients are needed to develop a reliable diagnostic test.

In this bi-center pilot study, we investigated whether breath gas analysis using multi-capillary column ion mobility spectrometry (MCC-IMS) could distinguish between symptomatic SARS-CoV-2 PCR-positive (PCR-positive) and -negative (PCR-negative) individuals. For this purpose, we performed breath gas measurements in individuals who presented to the ED with suspected COVID-19. RT-PCR was used as the gold standard to test for SARS-CoV-2, and breath gas measurements of the upper and lower respiratory tracts were performed in 380 participants (186 PCR positive and 194 PCR negative).

Material and methods

Materials

Material

Supplier

SpiroScout PC

Ganshorn, Niederlauer, Germany

PFT-Filter

Ganshorn, Niederlauer, Germany

ScoutTube

Ganshorn, Niederlauer, Germany

Adapter for PFT-Filter

Stromboli, Bochum, Germany

Connecting sample loop

KonMed, Rotkreuz, Switzerland

Incidin Plus 0.5%

Ecolab Deutschland GmbH, Monheim am Rhein, Germany

Software

VOCan software, B. Braun Melsungen AG, Center of Competence Breath Analysis, Dortmund, Germany.

VisualNow, Version 3.9.2 B. Braun Melsungen AG, Center of Competence Breath Analysis, Dortmund, Germany.

RapidMiner Version 9.2.001, Rapid, Boston, MA, USA.

SPSS V27, IBM, Armonk, NY, USA.

GraphPad Prism 7.04, GraphPad, La Jolla, CA, USA.

Study design and data acquisition

This case–control accuracy study was conducted at the University Hospital rechts der Isar, Technical University of Munich, Munich, Germany (TUM), and Charité-Universitätsmedizin Berlin, Campus Benjamin Franklin, Germany (CBF). A total of 396 adult (≥ 18 years) participants were recruited from patients presenting to the ED between September 23, 2020, and June 11, 2021 (TUM: 313, CBF: 83). No formal power calculation was performed at the outset of the study for the signals, and the effect sizes were yet to be determined.

The inclusion criteria were signs or symptoms of any respiratory system infection, fever, or radiological findings suggesting pulmonary manifestations of viral lung infection, and performance of reverse transcription-polymerase chain reaction (RT-PCR) testing for SARS-CoV-2. Equal proportions of patients with positive and negative test results were recruited for this study. Major exclusion criteria were inability to perform breath gas measurement, a history of SARS-CoV-2 infection, or participation in another clinical study prior to breath analysis, which could influence the results of the breath analysis according to the assessment of the principal investigator. Data obtained during routine clinical use were documented using an electronic Case Report Form (eCRF). Additionally, predefined clinical signs and symptoms (fever, shivering, cough, dyspnea, headache, limb pain, diarrhea, loss of smell or taste, and fatigue) and demographic and background variables (such as age, sex, and smoking habits) were documented.

Breath gas analysis was performed within 48 h of the RT-PCR testing. 120 min before the breath gas analysis, participants had to refrain from food, liquid intake (except water), or smoking. Written informed consent was obtained from each participant before inclusion in this study.

This study was approved by the Ethics Committee of the Technical University of Munich, School of Medicine, University Hospital rechts der Isar, Munich, Germany (approval no. 437/20 S-KH) and conducted in accordance with the Declaration of Helsinki. This study was registered at ClinicalTrials.gov (NCT04556318; Study-ID: HC-N-H-2004).

SARS-CoV-2 PCR

SARS-CoV-2 polymerase chain reaction (PCR) was performed using nasopharyngeal swabs. For the Munich site, PCR was performed by detection of the SARS-CoV-2 N gene using established routine diagnostic platforms at the Institute of Virology (Technical University of Munich), such as Cepheid GeneXpert (Sunnyvale, USA), Pathofinder RealAccurate (Maastricht, The Netherlands), or the QIAGEN NeuMoDx (Hilden, Germany) system, and Cycle threshold (Ct) values were available for every patient. For the Berlin site, the SARS-CoV-2 test on the Cobas® 6800 or 8800 system, Cepheid Xpress CoV-2 test, and Cobas® Liat® SARS-CoV-2 test were used.

Breath gas analysis by MCC-IMS

Breath gas samples for analysis using the MCC-IMS measurement tool were collected using an ultrasound-based spirometer (SpiroScout®, Ganshorn), which consisted of an ultrasound sensor with a disposable breathing tube used for breath sampling. The instrument was connected to a laptop computer with a dedicated analysis program for breath gas analysis (VOCan, B. Braun Melsungen AG, Center of Competence Breath Analysis, Dortmund, Germany). IMS allows the detection of gaseous volatile organic compounds in the trace range of ng/L to pg/L. As a result, gas samples with complex compositions and high moisture contents can be analyzed. The MCC-IMS system (Breath Discovery; B. Braun Melsungen AG, Center of Competence Breath Analysis) combines this highly sensitive method with gas chromatographic pre-separation using a multi-capillary column (1000 capillaries in parallel, inner diameter 40 µm, film thickness 200 nm, type OV-5; Multichrom Ltd., Moscow/Novosibirsk, Russia). A 95 MBq 63Ni β-radiation source was used to ionize the carrier gas (purified room air provided by REDMON, B. Braun Melsungen AG, Germany), which in turn ionized the sample via ion–molecule reactions. The ionized analytes were detected on a Faraday plate at the end of the drift tube by measuring the voltage. The data were displayed in a three-dimensional IMS chromatogram. Here, the peaks are defined by the drift time (via IMS), retention time (via MCC), and signal intensity, which indicates the relative concentration of the analyte.

Two breath samples were collected for each participant. In the first measurement, the breath of the oropharyngeal space was examined. Using SpiroScout, the air between 10 and 500 mL of each exhalation was passed over the sampling loop. Sampling was completed when the collection time added from the individual breaths was 10 s. Sample analysis was performed for 480 s. In the second measurement, the air exhaled from the lungs was analyzed. For this purpose, the sample was collected from the exhaled air, starting at a flow volume of 500 mL until flow reversal (inspiration). The collection time was 10 s, followed by sample analysis. This was followed by disinfection of the device with Incidin Plus 0.5%, a blank measurement to check the purity of the measuring device, and a room air measurement for comparison with the breath measurement.

Data analysis and statistics

Data analysis of breath gas measurements was performed independently of the RT-PCR results. Manual peak detection was performed using the VisualNow version 3.9 software (B. Braun Melsungen AG, Germany) [28]. The maximum peak values of the PCR-negative and PCR-positive groups and defined subgroups were compared using the Mann–Whitney U test.

Decision trees were generated using SPSS V27 [37]. The maximum values of all peaks considered for the decision trees were analyzed according to their diagnostic quality (area under the receiver operating characteristic curve). For each peak, the thresholds for 100% sensitivity and the threshold for 100% specificity were determined. True-positive (TP) and true-negative (TN) results detectable using these thresholds were counted, with a limit on the number of false positives (FPs) and false negatives (FNs) permitted. The peak with the largest sum of TP and TN results was selected (in the case of several peaks with the same sum, the peak with the lowest number in the given order was selected). The records that could be identified as TP and TN according to the selected peak were marked as “scored” and eliminated from the total database. For the remaining records, the procedure was repeated using a reduced database for all peaks that had not yet been considered. The procedure was terminated when no more than 20 records were identified as having TP or TN. The remaining records were classified as positive, with a setting sensitivity of 100% and a specificity of not less than 90% (in the case of 200 patients with negative PCR results). The algorithm was applied again to generate another decision tree using thresholds for lower sensitivity and specificity. The decision tree with the highest Youden index was selected for further consideration. Decision trees were generated for different subgroups based on site, sex, age, and cycle Ct values.

Principal component analysis (PCA) is an orthogonal linear transformation that transforms the given data into a new coordinate system to create a lower-dimensional subspace such that as little information as possible is lost and redundancy is reduced. PCA was used for dimensionality reduction by projecting maximal peak values onto only the first few principal components using SPSS functionality. The dimension was reduced by all components with eigenvalues > 1 in the transformation matrix, and the sum of their variances was 90% of the total variance. Additionally, linear discriminant analysis was performed by systematically selecting the peaks that provided the highest increase in accuracy in a step-by-step manner.

Statistical analyses were performed using SPSS V27 and GraphPad Prism software version 7.04. Statistical analyses were performed as indicated.

Results

From September 2020 to June 2021, 396 adult patients with signs of acute respiratory tract infection, fever, or radiological findings suggestive of viral lung infection were recruited from the EDs of TUM and CBF. Three hundred and eighty patients had evaluable breath gas data and PCR results (intention-to-treat population, ITT; Fig. 1). For 360 patients, no major overlaps or interference with disturbance variables was detected (per protocol population, PP).

Fig. 1
figure 1

Study profile. Patients were recruited between September 2020 and June 2021 at the Technical University of Munich (TUM) and Charité-Universitätsmedizin Berlin, Campus Benjamin Franklin, Germany (CBF); N Number of patients

Demographic and baseline clinical characteristics were relatively balanced between the PCR-positive and PCR-negative groups (Table 1). The mean age was 61.5, with more than half of the patients older than 60 years. The proportion of women was 38.7% (PCR positive) and 36.1% (PCR negative). The proportion of smokers was low overall but significantly higher in the PCR-positive group than in the PCR-negative group (p = 0.002, Fisher’s exact test). The median duration of symptoms at inclusion was 4.0 days (interquartile range [IQR], 2.0 to 7.0) in the PCR-positive group and 7.0 (IQR, 4.0 to 8.5) in the PCR-negative group.

Table 1 Patient characteristics at baseline

Cough and fever were the most common symptoms in the PCR-positive group; however, these complaints were also very common in the PCR-negative group (Fig. 2). However, loss of smell or taste was almost exclusively reported in the PCR-positive group (fold difference 12.5 and 8.7, respectively).

Fig. 2
figure 2

Symptoms. Symptoms are listed in descending order of differential frequency in PCR-positive and -negative patients. It is noted that patients could report more than one symptom. Numbers indicate fold difference between SARS-CoV-2 PCR positive and negative. * significant (p < 0.05). Significance was assessed by Fisher’s exact test

Peripheral oxygen saturation on room air (SpO2) and laboratory values of C-reactive protein (CRP), lactate dehydrogenase (LDH), D-dimer, and absolute lymphocyte count were not significantly different between the SARS-CoV-2 PCR-positive and -negative groups (Table 2). Only procalcitonin (PCT), a marker of bacterial infection, was significantly higher in the SARS-CoV-2 PCR-negative group.

Table 2 Laboratory tests

Ct values, which are semi-quantitative values of the concentration of viral genetic material in a patient’s sample as determined by RT-PCR, were available for the positive samples at the TUM site [38]. Almost two-thirds of the participants had a Ct value < 30, which is generally considered an infectious viral load (Additional file 1: Fig. S1) [39].

Breath gas analysis

Peak intensity

Breath gas analysis using the IMS measurement tool was performed within 48 h of SARS-CoV-2 RT-PCR testing. For all patients, samples were collected from the oropharyngeal space (throat, T) and lungs (L) (for details, see “Breath gas analysis by MCC-IMS” section).

In the IMS chromatograms, 110 peaks were identified by manual peak selection. Overall, 54 peaks (11 peaks for ITT-L only, 8 for ITT-T only, and 35 for both ITT-L and ITT-T) showed statistically significant differences in peak intensity between the SARS-CoV-2 PCR-negative and -positive groups in the ITT population (Additional file 1: Table S1). Some maxima of peaks showed significant differences in peak intensity between SARS-CoV-2 PCR positive and PCR negative for both throat and lung samples, with similar peak intensities for both study sites (Table 3, Peak 1). For other peaks, differences between sites were observed, although consistent and significant differences between SARS-CoV-2 PCR-positive and PCR-negative samples were still evident (Table 3, Peak 2).

Table 3 Exemplary characterization of two maxima of peaks in both lung and throat samples

Decision trees

A commonly used tool in operations research is decision trees, which were also applied in our previous preclinical project (for details, see “Data analysis and statistics” section). For the PP and ITT groups, the sensitivity and specificity were well over 80% for the lung, throat, and combined samples (Table 4).

Table 4 Differentiation of breath gas samples by decision tree analysis

We wondered whether the sensitivity of detecting SARS-CoV-2 infection depended on the participant’s viral load. However, there was no trend toward better sensitivity in the group with a Ct value of less than 30 (higher viral load) compared to the groups with a Ct value of 30–35 and a Ct value greater than 35 (Additional file 1: Table S2). However, subgroup analyses implied an influence of site, sex, and age. Combining these influencing variables would reduce the available number of cases to such an extent that decision trees would no longer be meaningful. In addition, despite the relatively high values for sensitivity and specificity, decision trees did not prove robust to changes in the database (e.g., the use of only a part of the dataset).

Linear discriminant and principal component analyses

Next, we evaluated the potential of model building using linear discriminant analysis. We selected 11 peaks with the smallest Area under the ROC Curve (AUC(ROC)) to differentiate between SARS-CoV-2 PCR-negative and PCR-positive samples from the ITT-L group. For each of these peaks, the comparison between the maximum peak of the SARS-CoV-2 PCR-positive and PCR-negative groups was statistically significant (p < 0.001). Linear discriminant analysis classified the two groups with a sensitivity of 74% and specificity of 52% (Table 5).

Table 5 Results of linear discriminant analyses and systematic searches for the ITT-L group

Dimensional reduction was attempted using PCA (“Data Analysis and statistics” section). Only three of the 11 peaks mentioned previously had an eigenvector with an eigenvalue greater than 1. With the help of these three factors, a cumulative variance of 78% can be explained. However, the dimension reduction (from 11 to 3) led to a worse classification accuracy than that of the discriminant analysis with the 11 original peak maxima (Additional file 1: Table S3). An extension of the most selective combinations by the single peaks with the highest increase in accuracy was found only by a systematic search, resulting in a sensitivity of 77% and a specificity of 67% (Table 5). Similar results were obtained when only TUM site data were examined. The systematic search for CBF data only revealed a significantly better classification, again stressing the influence of the site (Table 5).

Discussion

Respiratory tract infections can cause metabolic changes that lead to alterations in the respiratory VOC profile, which enables their potential use in non-invasive breath gas diagnostics. In a preclinical study, we have previously shown the reliable differentiation of SARS-CoV-2 from other respiratory viruses using MCC-IMS analysis of air collected from virus-infected cell cultures [28]. The first clinical studies on breath analysis using gas chromatography–mass spectrometry revealed high sensitivity for detecting SARS-CoV-2 but used only small sample sizes (reviewed in [40]). A Dutch study of 4510 non-hospitalized participants also found a high sensitivity for SARS-CoV-2 detection by breath gas analysis with eNose, which uses pattern recognition in cross-reactive metal oxide semiconductor sensors [36]. Using similar technology, Nurputra et al. reported high accuracy in detecting SARS-CoV-2 in samples taken from 43 COVID-19-positive and 40 -negative patients [41].

In our study, 380 participants underwent breath gas analysis using MCC-IMS in both lung and throat spaces. The inclusion criteria were defined as comparable to the established test routine for suspected SARS-CoV-2 infection: symptoms of any respiratory system infection, fever, or radiological findings for viral pneumonia to achieve high consistency with real-world clinical practice.

Demographic and baseline clinical characteristics were overall relatively similar between the PCR-positive and PCR-negative groups. The proportion of smokers was significantly higher in the PCR-positive group. However, considering the overall low smoking rate, the selection bias can be considered to be low. Also, the duration of symptoms upon inclusion was significantly shorter in the PCR-positive group when compared to the PCR-negative group. As this study was conducted as a pilot, measurements were limited to a single time point. The duration of symptoms at the time of inclusion varied widely and showed considerable overlap between the two groups, with interquartile ranges (IQR) of 2–7 days for the PCR-positive group and 4–8.5 days for the PCR-negative group. The VOC profile probably undergoes certain changes during the course of the infection. To gain a more precise understanding of these changes, longitudinal measurements throughout the entire duration of the infection would be desirable. However, for clinical utility, it is crucial that a specific VOC profile remains detectable over an extended period throughout the infection.

Furthermore, consistent with the literature, loss of smell or taste was almost exclusively reported in the PCR-positive group [42]. There were no significant differences in laboratory values between the two groups, except for PCT, which was higher in the PCR-negative group, as expected, due to the probably higher incidence of bacterial infections.

Breath gas analysis using MCC-IMS identified 54 peaks with a statistically significant difference in peak intensity between the PCR-negative and PCR-positive groups. However, finding a classification system that allows reliable differentiation between the two groups is challenging. An algorithm based on decision trees allowed the detection of SARS-CoV-2 with a sensitivity and specificity well above 80%. Subgroup analyses implied the influence of the study site, sex, and age. However, when the generation of decision trees was restricted to a selection of the available datasets, the transferability to the remaining datasets was limited. Furthermore, an influence of the viral load on sensitivity could not be shown. A major cause could be that, despite statistically significant differences in the peak values of SARS-CoV-2 PCR negative and PCR positive, only a few values can be correctly assigned based on a threshold value without simultaneously obtaining incorrect assignments.

Next, we applied a model-building approach using a linear discriminant analysis. In this analysis, only modest values for sensitivity (73.7%) and specificity (51.5%) were obtained. Applying PCA in combination with a systematic search showed a sensitivity of 77.4% and a specificity of 67.0% for SARS-CoV-2-PCR positivity. However, the results depend on the current database and can therefore be assumed not to be transferable to other datasets. Furthermore, it is unclear how a suitable peak can be selected using an algorithm.

As another analysis method, we explored the use of artificial intelligence to create a classification system. However, training computer vision machine learning models (ResNet18, Convolutional Neural Network [43, 44]) on breath gas data failed. Possible reasons for this were the low sample numbers for this initially unplanned analysis and the data points containing little discriminative information from a computer vision perspective.

A sufficient sample size is always a challenge for model building. The inclusion of 380 participants, 186 of whom were PCR positive, in an emergency room setting is remarkable, but especially for a machine learning approach, much higher patient numbers would have been necessary. Highly sensitive breath gas analysis is susceptible to environmental factors such as indoor air or disinfectant used, which became evident in our data because of the influence of the site on peak intensity. This highlights the need for multiple sites to develop a robust algorithm that is not overfitted for one site, and demonstrates the limited transferability of other monocentric breath gas studies.

During the recruitment period of this study, the ancestral variant initially prevailed, but in the first half of 2021, it was increasingly replaced by the Alpha variant (B.1.1.7.) [45]. The concept of breath gas analysis is based on the idea that metabolic changes create specific VOC profiles depending on the pathogen involved (as described in [26, 27]). It would be plausible to assume that different variants of SARS-CoV-2 may potentially generate slightly different VOC profiles. However, it is also reasonable to speculate that an association with the species SARS-CoV-2 should still be possible. Calibration to the currently predominant variant may be advisable to increase the accuracy of the test.

The present study distinguishes itself from other similar projects due to its significantly larger sample size and the inclusion of two study sites (see [31,32,33]). Moreover, we regard the apparent selection bias, the recruitment of symptomatic individuals from the emergency department, to be a strength of the study as it represents a highly medically relevant group. These factors enhance the statistical power, generalizability, and clinical significance of the findings, making it a valuable contribution to the field of breath gas analysis for SARS-CoV-2 detection.

Conclusion

In this bi-center clinical pilot study, we demonstrate that breath gas analysis with MCC-IMS allows the detection of SARS-CoV-2 infection in patients with a sensitivity and specificity of more than 80% in a real-life ED setting. However, the algorithms used in the analysis depend on the current database, which makes the transferability of the results to other datasets difficult. Despite the limited robustness of the algorithms developed in this study, the large number of significantly different peaks between the PCR-positive and PCR-negative groups seems promising. The use of drift tubes adapted to the now-known distribution spectrum of the peaks could lead to a much better separation of the frequently superimposed peaks in this dataset. In addition, structural modifications of the measuring device to improve its disinfect ability could reduce the disturbing influence of disinfectants.

This study showed the general feasibility of such a non-invasive point-of-care diagnostic test without the need for a high number of consumables. SARS-CoV-2 has highlighted the role of rapid diagnostic testing in pandemic control. Further research on technical advancements, analysis of highly complex datasets and evaluation in a multicenter environment are required to make breath gas analysis a valuable tool in the future.

Availability of data and materials

Written informed consent was obtained from all participants before inclusion in this study. The datasets generated and analyzed during the current study are available from the co-author Heike Fries (email: heike.fries@bbraun.com) upon reasonable request.

Abbreviations

AMS:

Antimicrobial stewardship

AUC(ROC):

Area under the ROC Curve

CBF:

Charité: Universitätsmedizin Berlin, Campus Benjamin Franklin, Germany

COVID-19:

Coronavirus disease 2019

CRP:

C-reactive protein

Ct:

Cycle threshold

eCRF:

Electronic Case Report Form

ED:

Emergency department

FN:

False negative

FP:

False positive

IMS:

Ion mobility spectrometry

IQR:

Interquartile range

ITT:

Intention-to-treat

L:

Lung

LDH:

Lactate dehydrogenase

MCC-IMS:

Multi-capillary column ion mobility spectrometry

MS:

Mass spectrometry

PCA:

Principal component analysis

PCR:

Polymerase chain reaction

PCT:

Procalcitonin

PP:

Per protocol

RT-PCR:

Real-time polymerase chain reaction

SARS-CoV-2:

Severe acute respiratory syndrome coronavirus 2

SD:

Standard deviation

SpO2:

Peripheral oxygen saturation on room air

T:

Throat

TAT:

Turnaround time

TN:

True negative

TP:

True positive

TUM:

University Hospital rechts der Isar, Technical University of Munich, Germany

VOC:

Volatile organic compound

References

  1. Johns Hopkins Coronavirus Resource Center. https://coronavirus.jhu.edu/. Accessed 15 Jan 2023.

  2. Forni G, Mantovani A, COVID-19 Commission of Accademia Nazionale dei Lincei R. COVID-19 vaccines: where we stand and challenges ahead. Cell Death Differ. 2021;28:626–39. https://doi.org/10.1038/s41418-020-00720-9.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  3. Fernandes Q, Inchakalody VP, Merhi M, Mestiri S, Taib N, Moustafa Abo El-Ella D, et al. Emerging COVID-19 variants and their impact on SARS-CoV-2 diagnosis, therapeutics and vaccines. Ann Med. 2022. https://doi.org/10.1080/07853890.2022.2031274.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Rickman HM, Rampling T, Shaw K, Martinez-Garcia G, Hail L, Coen P, et al. Nosocomial transmission of coronavirus disease 2019: a retrospective study of 66 hospital-acquired cases in a London teaching hospital. Clin Infect Dis. 2020. https://doi.org/10.1093/cid/ciaa816.

    Article  Google Scholar 

  5. Abbas M, Robalo Nunes T, Martischang R, Zingg W, Iten A, Pittet D, et al. Nosocomial transmission and outbreaks of coronavirus disease 2019: the need to protect both patients and healthcare workers. Antimicrob Resist Infect Control. 2021. https://doi.org/10.1186/s13756-020-00875-7.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Crotty MP, Meyers S, Hampton N, Bledsoe S, Ritchie DJ, Buller RS, et al. Impact of antibacterials on subsequent resistance and clinical outcomes in adult patients with viral pneumonia: an opportunity for stewardship. Crit Care. 2015. https://doi.org/10.1186/s13054-015-1120-5.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Barlam TF, Cosgrove SE, Abbo LM, Macdougall C, Schuetz AN, Septimus EJ, et al. Implementing an antibiotic stewardship program: guidelines by the Infectious Diseases Society of America and the Society for Healthcare Epidemiology of America. Clin Infect Dis. 2016. https://doi.org/10.1093/cid/ciw118.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Dinnes J, Deeks JJ, Adriano A, Berhane S, Davenport C, Dittrich S, et al. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst Rev. 2020. https://doi.org/10.1002/14651858.CD013705.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Vandenberg O, Martiny D, Rochas O, van Belkum A, Kozlakidis Z. Considerations for diagnostic COVID-19 tests. Nat Rev Microbiol. 2021. https://doi.org/10.1038/s41579-020-00461-z.

    Article  PubMed  Google Scholar 

  10. Diel R, Nienhaus A. Point-of-care COVID-19 antigen testing in German emergency rooms—a cost-benefit analysis. Pulmonology. 2022;28:164–72. https://doi.org/10.1016/j.pulmoe.2021.06.009.

    Article  PubMed  CAS  Google Scholar 

  11. Osterman A, Baldauf H-M, Eletreby M, Wettengel JM, Afridi SQ, Fuchs T, et al. Evaluation of two rapid antigen tests to detect SARS-CoV-2 in a hospital setting. Med Microbiol Immunol. 2021. https://doi.org/10.1007/s00430-020-00698-8.

    Article  PubMed  PubMed Central  Google Scholar 

  12. World Health Organization (WHO). Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immunoassays. Interim Guid. 2020. p. 9. https://www.who.int/publications/i/item/antigen-detection-in-the-diagnosis-of-sars-cov-2infection-using-rapid-immunoassays. Accessed 20 Nov 2020.

  13. Paul-Ehrlich-Institut. Minimum criteria for rapid SARS-CoV-2 antigen tests pursuant to section 1 para 1 sentence 1 TestVO (statutory test regulation): rapid antigen tests (a) performance indicators. Paul-Ehrlich-Institut. 2021;2–4.

  14. Del Vecchio C, Cracknell Daniels B, Brancaccio G, Brazzale AR, Lavezzo E, Ciavarella C, et al. Impact of antigen test target failure and testing strategies on the transmission of SARS-CoV-2 variants. Nat Commun. 2022;13:5870. https://doi.org/10.1038/s41467-022-33460-0.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  15. Möckel M, Corman VM, Stegemann MS, Hofmann J, Stein A, Jones TC, et al. SARS-CoV-2 antigen rapid immunoassay for diagnosis of COVID-19 in the emergency department. Biomarkers. 2021;26:213–20. https://doi.org/10.1080/1354750X.2021.1876769.

    Article  PubMed  Google Scholar 

  16. Yang SM, Lv S, Zhang W, Cui Y. Microfluidic point-of-care (POC) devices in early diagnosis: a review of opportunities and challenges. Sensors. 2022. https://doi.org/10.3390/s22041620.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Yin B, Wan X, Sohan ASMMF, Lin X. Microfluidics-based POCT for SARS-CoV-2 diagnostics. Micromachines. 2022. https://doi.org/10.3390/mi13081238.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Walker HJ, Burrell MM. Could breath analysis by MS could be a solution to rapid, non-invasive testing for COVID-19? Bioanalysis. 2020. https://doi.org/10.4155/bio-2020-0125.

    Article  PubMed  Google Scholar 

  19. Cumeras R, Figueras E, Davis CE, Baumbach JI, Gràcia I. Review on ion mobility spectrometry. Part 1: current instrumentation. Analyst. 2015. https://doi.org/10.1039/c4an01100g.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Cumeras R, Figueras E, Davis CE, Baumbach JI, Gràcia I. Review on ion mobility spectrometry. Part 2: hyphenated methods and effects of experimental parameters. Analyst. 2015. https://doi.org/10.1039/c4an01101e.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Vautz W, Nolte J, Fobbe R, Baumbach JI. Breath analysis—Performance and potential of ion mobility spectrometry. J Breath Res. 2009. https://doi.org/10.1088/1752-7155/3/3/036004.

    Article  PubMed  Google Scholar 

  22. Amann A, Miekisch W, Schubert J, Buszewski B, Ligor T, Jezierski T, et al. Analysis of exhaled breath for disease detection. Annu Rev Anal Chem. 2014. https://doi.org/10.1146/annurev-anchem-071213-020043.

    Article  Google Scholar 

  23. Schivo M, Aksenov AA, Linderholm AL, McCartney MM, Simmons J, Harper RW, et al. Volatile emanations from in vitro airway cells infected with human rhinovirus. J Breath Res. 2014. https://doi.org/10.1088/1752-7155/8/3/037110.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Aksenov AA, Sandrock CE, Zhao W, Sankaran S, Schivo M, Harper R, et al. Cellular scent of influenza virus infection. ChemBioChem. 2014;15:1040–8. https://doi.org/10.1002/cbic.201300695.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  25. Purcaro G, Rees CA, Wieland-Alter WF, Schneider MJ, Wang X, Stefanuto PH, et al. Volatile fingerprinting of human respiratory viruses from cell culture. J Breath Res. 2018. https://doi.org/10.1088/1752-7163/aa9eef.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Bos LDJ, Sterk PJ, Schultz MJ. Volatile metabolites of pathogens: a systematic review. PLoS Pathog. 2013. https://doi.org/10.1371/journal.ppat.1003311.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Gould O, Ratcliffe N, Król E, De Lacy Costello B. Breath analysis for detection of viral infection, the current position of the field. J Breath Res. 2020. https://doi.org/10.1088/1752-7163/ab9c32.

    Article  PubMed  Google Scholar 

  28. Feuerherd M, Sippel AK, Erber J, Baumbach JI, Schmid RM, Protzer U, et al. A proof of concept study for the differentiation of SARS-CoV-2, hCoV-NL63, and IAV-H1N1 in vitro cultures using ion mobility spectrometry. Sci Rep. 2021. https://doi.org/10.1038/s41598-021-99742-7.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Chen H, Qi X, Ma J, Zhang C, Feng H, Yao M. Breath-borne VOC biomarkers for COVID-19. medRxiv. 2020. https://doi.org/10.1101/2020.06.21.20136523.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Ruszkiewicz DM, Sanders D, O’Brien R, Hempel F, Reed MJ, Riepe AC, et al. Diagnosis of COVID-19 by analysis of breath with gas chromatography-ion mobility spectrometry—a feasibility study. EClinicalMedicine. 2020. https://doi.org/10.1016/j.eclinm.2020.100609.

    Article  PubMed  Google Scholar 

  31. Steppert C, Steppert I, Sterlacci W, Bollinger T. Rapid detection of SARS-CoV-2 infection by multicapillary column coupled ion mobility spectrometry (MCC-IMS) of breath. A proof of concept study. J Breath Res. 2021. https://doi.org/10.1088/1752-7163/abe5ca.

    Article  PubMed  Google Scholar 

  32. Ibrahim W, Cordell RL, Wilde MJ, Richardson M, Carr L, Sundari Devi Dasi A, et al. Diagnosis of COVID-19 by exhaled breath analysis using gas chromatography–mass spectrometry. ERJ Open Res. 2021. https://doi.org/10.1183/23120541.00139-2021.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Grassin-Delyle S, Roquencourt C, Moine P, Saffroy G, Carn S, Heming N, et al. Metabolomics of exhaled breath in critically ill COVID-19 patients: a pilot study. EBioMedicine. 2021. https://doi.org/10.1016/j.ebiom.2020.103154.

    Article  PubMed  Google Scholar 

  34. Shan B, Broza YY, Li W, Wang Y, Wu S, Liu Z, et al. Multiplexed nanomaterial-based sensor array for detection of COVID-19 in exhaled breath. ACS Nano. 2020;14:12125–32. https://doi.org/10.1021/acsnano.0c05657.

    Article  PubMed  CAS  Google Scholar 

  35. Wintjens AGWE, Hintzen KFH, Engelen SME, Lubbers T, Savelkoul PHM, Wesseling G, et al. Applying the electronic nose for pre-operative SARS-CoV-2 screening. Surg Endosc. 2020. https://doi.org/10.1007/s00464-020-08169-0.

    Article  PubMed  PubMed Central  Google Scholar 

  36. de Vries R, Vigeveno RM, Mulder S, Farzan N, Vintges DR, Goeman JJ, et al. Ruling out SARS-CoV-2 infeciton using exhaled breath analysis by electronic nose in a public health setting. medRxiv. 2021. https://doi.org/10.1164/ajrccm-conference.2021.203.1_meetingabstracts.a3768.

    Article  Google Scholar 

  37. Baizyldayeva UB, Uskenbayeva RK, Amanzholova ST. Decision making procedure: applications of IBM SPSS cluster analysis and decision tree. World Appl Sci J. 2013;21:1207–12. https://doi.org/10.5829/idosi.wasj.2013.21.8.2913.

    Article  Google Scholar 

  38. Public Health England. Understanding cycle threshold (Ct) in SARS-CoV-2 RT-PCR: a guide for health protection teams. Public Heal Engl; 2020. p. 1–12.

    Google Scholar 

  39. RKI—Coronavirus SARS-CoV-2—Empfehlungen zu Isolierung und Quarantäne bei SARS-CoV-2-Infektion und -Exposition, Stand 2.5.2022. https://www.rki.de/DE/Content/InfAZ/N/Neuartiges_Coronavirus/Quarantaene/Absonderung.html;jsessionid=1FC86EBE9E72AA0888DD33D7C7E5F806.internet082?nn=13490888. Accessed 26 May 2022.

  40. Subali AD, Wiyono L, Yusuf M, Zaky MFA. The potential of volatile organic compounds-based breath analysis for COVID-19 screening: a systematic review & meta-analysis. Diagn Microbiol Infect Dis. 2022. https://doi.org/10.1016/j.diagmicrobio.2021.115589.

    Article  PubMed  Google Scholar 

  41. Nurputra DK, Kusumaatmaja A, Hakim MS, Hidayat SN, Julian T, Sumanto B, et al. Fast and noninvasive electronic nose for sniffing out COVID-19 based on exhaled breath-print recognition. NPJ Digit Med. 2022;5:1–17. https://doi.org/10.1038/s41746-022-00661-2.

    Article  Google Scholar 

  42. Lee Y, Min P, Lee S, Kim SW. Prevalence and duration of acute loss of smell or taste in COVID-19 patients. J Korean Med Sci. 2020. https://doi.org/10.3346/JKMS.2020.35.E174.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Hosny A, Parmar C, Quackenbush J, Schwartz LH, Aerts HJWL. Artificial intelligence in radiology. Nat Rev Cancer. 2018. https://doi.org/10.1038/s41568-018-0016-5.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Kaissis G, Ziller A, Passerat-Palmbach J, Ryffel T, Usynin D, Trask A, et al. End-to-end privacy preserving deep learning on multi-institutional medical imaging. Nat Mach Intell. 2021;3:473–84. https://doi.org/10.1038/s42256-021-00337-8.

    Article  Google Scholar 

  45. Robert Koch Institut. Bericht zu Virusvarianten von SARS-CoV-2 in Deutschland. Robert Koch Inst. 2021;2:1–19.

    Google Scholar 

Download references

Acknowledgements

The authors would like to acknowledge the use of AI language models, specifically ChatGPT and DeepL, for their assistance in translation and grammar checks during the preparation of this manuscript. They would also like to express their gratitude to Editage for their professional language editing services, which helped improve the overall clarity and readability of the manuscript. While these tools and editing services have provided valuable support, they would like to emphasize that the responsibility for the content and accuracy of the manuscript rests with the authors.

Funding

Open Access funding enabled and organized by Projekt DEAL. This project received funding from B. Braun Melsungen AG (sponsor).

Author information

Authors and Affiliations

Authors

Contributions

FV, HF, and CS conceived and designed the study. FV, JE, WB, ED, GS, and LJ recruited participants and/or collected data. NB performed the statistical analysis of the data. MF, JS, ED, and FG contributed to the data interpretation. FV and WB wrote the manuscript. RMS, UP, RS, and CDS supervised and administered the study protocols. All the authors reviewed the final version of the manuscript for approval and concurred with the decision to submit the manuscript for publication.

Corresponding author

Correspondence to Florian Voit.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in this study involving human participants were in accordance with the ethical standards of the institutional and national research committee and the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. This study was approved by the Ethics Committee of the Technical University of Munich, School of Medicine, University Hospital rechts der Isar, Munich, Germany (approval no. 437/20 S-KH) and conducted in accordance with the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

FV reports personal fees, non-financial support from Gilead Sciences, and grants from MSD outside the submitted work. The institution of E.D.W., R.S., and W.B. received Funding from Inflammatix, Gentian, and B. Braun Melsungen. JS reports grants, personal fees, and non-financial support from AbbVie, Gilead, Janssen, and Dr. Falk GmbH. UP received personal fees from Abbott, Abbvie, Arbutus, Gilead, GSK, J&J, MSD, Roche, Sanofi, Sobi, and Vaccitech. UP is co-founder and shareholder of SCG Cell Therapy. The UP laboratory received grants from Hoenle AG and SCG Cell Therapy. The State of Bavaria supported the project via research networks FOR-COVID and Bay-VOC, by the project “Virological and immunological determinants of COVID-19 pathogenesis—lessons to get prepared for future pandemics (KA1-Co-02 “COVIPA”),” a grant from the Helmholtz Association's Initiative and Networking Fund, by the European Commission FET Open Grant VIROFIGHT (Grant No. 899619). CS reports grants and personal fees from AstraZeneca; personal fees and non-financial support from B. Braun Melsungen; personal fees from BioNTech; grants, personal fees, and non-financial support from Gilead Sciences; grants and personal fees from Janssen-Cilag; personal fees from Eli Lilly; personal fees from Formycon; personal fees from Pfizer; personal fees from Roche; other from Apeiron; grants and personal fees from MSD; grants from Cepheid; personal fees from GSK; personal fees from molecular partners; other from Eli Lilly; personal fees from SOBI during the conduct of the study; personal fees from AbbVie; personal fees from MSD; personal fees from Synairgen; and grants and personal fees from ViiV Healthcare outside the submitted work.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Table S1.

Significantly different peaks between the PCR-positive and PCR-negative groups for lung and throat samples of the ITT population. Table S2. Sensitivity of SARS-CoV-2 detection in decision tree analysis for different groups of Ct values. Table S3. Performance of discriminant analysis for reduction of dimensions via principal component analysis (ITT-L). Fig. S1. SARS-CoV-2 PCR was performed as described in “Materials and methods” section. Ct values were available for the Munich site only. A lower Ct value indicates higher concentrations of the detected virus genome.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Voit, F., Erber, J., Feuerherd, M. et al. Rapid point-of-care detection of SARS-CoV-2 infection in exhaled breath using ion mobility spectrometry: a pilot study. Eur J Med Res 28, 318 (2023). https://doi.org/10.1186/s40001-023-01284-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40001-023-01284-3

Keywords